HARPS ThAr Lamps Monitoring
ThAr exposures
Following is the plot of the time evolution of the intensity measured
on the two fibers from the two ThAr calibration lamps.
"Fiber A" is illuminated by "lamp4" (or "ThAr1"),
"Fiber B" is illuminated by "lamp5" (or "ThAr2").
The data are obtained from the "spot" files output
by the pipeline for every ThAr calibration exposure.
Only the last exposure of the day, taken just before
the beginning of the night, is considered.
A trend different from flat might be due to a lamp change, a lamp aging,
a change in the injection (alignment), or in the fibers (stress).
For diagnostic these plots have to be compared to the
ThAr lamps history, the
flat fields and the day sky exposures.