[page auto-refreshes after 300 sec]            [press Ctrl+R to enforce refresh of scores and dates]
CRIRES trending system: HEALTH CHECK plot
Last update: 2010-03-18T04:47:50 (UT)
| now: 2010-03-18T06:15:33 (UT)
HELP USERS-GUIDE MORE
same group: flat Hal_IR   ThAr
no OPSLOG data
space
 
N_data:
close window history: 2009   2010   HEALTH
 
*Date on this monitor changes at 21:00 UT
Print: Information and research: News:
- png file - advanced studies: QC1 browser | QC1 plotter
 
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB no MEDIAN 407.24203 ADU/sec VAL150,3000 112 lamp_eff this | last_yr | all lamp efficiency (flux) on chip 2, WREF=1090.4
2 QC1DB no MEDIAN 144.88257 ADU/sec VAL50,3000 112 lamp_eff this | last_yr | all lamp efficiency (flux) on chip 3, WREF=1090.4
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
This plot

Lamp efficiency: flux of ThAr lamp.

The lamp efficiency for the ThAr lamp is monitored using the daily health check setting with reference wavelength (on chip 3) of 1090.4nm. It gives the average flux of lines that could be identified with the ThAr catalog. This value should be within the indicated thresholds for each chip. As there is strong vignetting and contamination from adjacent orders on chips 1 and 4, only the results for chips 2 and 3 are plotted.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up