| zero points of photometric nights |
| 2006-09-03 to
2006-12-03 |
|
CHIP1 |
CHIP2 |
| Filter |
zero point |
range |
frames |
zero point |
range |
frames |
| U | 24.15 +/- 0.05
| 24.08 ... 24.26 | 16 | 24.25 +/- 0.07
| 24.10 ... 24.35 | 12 |
| B | 27.62 +/- 0.04
| 27.58 ... 27.71 | 16 | 27.63 +/- 0.05
| 27.59 ... 27.77 | 14 |
| V | 28.10 +/- 0.05
| 27.96 ... 28.14 | 17 | 28.12 +/- 0.03
| 28.04 ... 28.16 | 14 |
| R | 28.35 +/- 0.07
| 28.15 ... 28.42 | 17 | 28.35 +/- 0.02
| 28.33 ... 28.40 | 14 |
| I | 27.69 +/- 0.04
| 27.55 ... 27.72 | 18 | 27.71 +/- 0.03
| 27.64 ... 27.76 | 16 |
avg. zero point average of the zero points per frame of all photometric
nights, weighted by their inverse error
range range of zero points observed in the photometric nights
frames total number of frames from photometric nights
For daily FORS2 health check, focus on:
boxes per standard filter (BVI_BESSELL, UR_Special): The DFO generated
zeropoints (black) come per frame, with errorbars. Blue are the values generated
by the quick-look pipeline on Paranal, one value per pointing. There may
be several entries per night.
Values are generated for both CHIP1 (upper row) and CHIP2,
2x2 binning only. Blue points come from pipeline log files
transferred daily by PSO. Other points are from regular QC processing
at Garching.
|