[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
FORS2 trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-05-04T16:11:42 (UT)
  | now: 2024-05-04T16:48:26 (UT) muc01 QC pipeline: fors-5.6.10 (installed 2024-04-19)
same group: SCR_FF_RI SCR_FF_uv SCR_FF_b   SCR_FF_RMS_RI SCR_FF_RMS_uv SCR_FF_RMS_b
close window history: 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023   2024   HEALTH
  ... ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 1.38e+03 ADU 60%± 831 23 sigma_master this | last_yr | all screen flat rms in ADU for b_HIGH, low gain, chip 1
2 QC1DB MEDIAN 1.47e+03 ADU 60%± 882 23 sigma_master this | last_yr | all screen flat rms in ADU for b_HIGH, low gain, chip 2
3 QC1DB MEDIAN 1.25e+03 ADU 60%± 748 24 sigma_master this | last_yr | all screen flat rms in ADU for b_HIGH, high gain, chip 1
4 QC1DB MEDIAN 1.32e+03 ADU 60%± 792 24 sigma_master this | last_yr | all screen flat rms in ADU for b_HIGH, high gain, chip 2
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 3|
data source:fors2_scrflat
(QC1 database)
dataset:sigma_master
median:1.38e+03ADU
rel. thresholds: ±60%553.707324...2214.829296ADU
N_data plotted:23
[click plot for closeup]
Plot 2
scores: 3|
data source:fors2_scrflat
(QC1 database)
dataset:sigma_master
median:1.47e+03ADU
rel. thresholds: ±60%587.817676...2351.270704ADU
N_data plotted:23
[click plot for closeup]
Plot 3
scores: 2|
data source:fors2_scrflat
(QC1 database)
dataset:sigma_master
median:1.25e+03ADU
rel. thresholds: ±60%498.957836...1995.831344ADU
N_data plotted:24
[click plot for closeup]
Plot 4
scores: 2|
data source:fors2_scrflat
(QC1 database)
dataset:sigma_master
median:1.32e+03ADU
rel. thresholds: ±60%527.95559...2111.82236ADU
N_data plotted:24
[click plot for closeup]
This plot

The FlatBlue+1 and FlatBlue+2 lamps are used for imaging screen flats.
The RMS is calculated from the master screen flat image pixel sample (2x2 binning) X: @200 @1800 and Y: @400 @800 covering the unvignetted area of chip1 and chip2.

DB Name: fors2_scrflat
TPL.ID: FORS2_img_cal_scrflat
QC1 DB parameter: sigma_master

 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up