This is the Health Check plot to monitor the efficiency of the GIRAFFE arclamp. We display the QC1 parameter lamp_effic which is the average (mean intensity/EXPTIME) flux of all emission lines, in ADU/sec, for all fibres except for the SIMCAL fibres.

This plot monitors the HC data taken in high resolution (H525.8B, all slits, box 1-5) and low resolution (L543.1, Medusa1 and 2 only, boxes 6 and 7).

If the intensity drops in one of the fibre systems, there is most likely a problem with that fibre system. If the fibre systems all drop simultaneously, this is likely to be a problem with the calibration lamp itself.

The quality of the arclamp exposures is quite critical for the quality of the dispersion solution. An intensity drop increases the rms of the dispersion solution significantly (see rms plot under "wavelength calibration", around epoch 2007-10).