This is the FULL trending plot for the monitoring of the GIRAFFE CCD linearity and gain. These parameters are derived from image flats. These data are taken about monthly. Since 2008-07, they are acquired with the common detector monitoring template and reduced with the corresponding common recipe.

The CCD was upgraded between 2008-03 and 2008-05. Data from the new CCD are marked.

A sequence of image flats is exposed between 0.5 and 220 secs. A third-order polynomial is fitted to the mean exposure level as a function of exposure time. An effective non-linearity correction is derived from this and plotted in box 1. This is available only since 2008.

The gain (in electrons per ADU) is also monitored here, as box 2. It is measured from the difference between two identical raw frames, by comparing the square root of the signal to its measured rms. Plotted here is the asymptotic value for exptime → 0 which is the true gain. Find the complete correlation as plot no. 4 here. The value stored in the file headers is very close to the one plotted here.