This is the trending plot for the proper localization of fibres in the GIRAFFE flat field calibration data. We display the QC1 parameter num_fib (PRO.SLIT.NFIBRES). The purpose of this plot is to detect fibres with missing signal (broken fibres, un-illuminated ones), or pipeline failures.

The plots are arranged per fibre system. Within each fibre system, all settings are displayed. Their num_fib parameter should be almost the same (see below). Hence both the regular Health Check calibration, and the science-driven daytime calibrations are included here.

The IFU settings show some small variation around 315 since some of the last fibres may be truncated, depending on the setting. Most likely, problems will show up only per fibre bundle (sets of 20 fibres).

A detailed record of the performance of each fibre, and a record of the available/missing fibres, is available here.

You may want to check a raw file visually against the reference raw file displays.