This is the instrument performance plot for the GIRAFFE fibre system. There are five different fibre systems: Medusa1/2, IFU1/2, and Argus. For each of them, we display the QC1 parameter num_fib (PRO.SLIT.NFIBRES), in order to monitor the number of available fibres. A sudden deviation from the nominal behaviour might indicate issues with broken fibres or fibre illumination problems. A certain minor deviation from a previous number might also be due to a slightly moving spectral format. The pipeline extracts only fibre signals not truncated by the CCD edge.
The plots are arranged per fibre system. They are fed from the regular Health Check calibrations, taken as L543.1 setting every 3rd day in the Medusa slits, and once a week in the IFU and Argus slits.
The reference values indicated here are the maximum values observed in the history of the instrument.