This is the instrument performance plot for the GIRAFFE resolution. There are five different fibre systems: Medusa1/2, IFU1/2, and Argus, and two different gratings, HR and LR. For the HR grating, we display the resolving power R = λ0/Δλ. Δλ is the QC1 parameter mean_resol (QC.RESOL.MEAN) which is the mean of all accepted emission line widths. HC data are taken every 3 days for the Medusa slits, and every 7 days for the IFU/Argus slits. The monitoring is done in the H525.8B setting for the HR grating and L543.1 for the LR grating. For the LR grating, only the two Medusa settings are taken.

The reference values are marked in green.

The HR values for the first years (until 2007), when calculated with the then available pipeline versions, were systematically wrong due to a bug. Some of these values have been reprocessed with the current pipeline (giraf-2.5 and higher), they are displayed as separate datasets as indicated in plots #1, #2, #6, and #7. They confirm that the offset in resolution observed for the HR grating was not real. The observed degradation in the HR resolution until 2015 was real. It was successfully remedied by an intervention, with the values being stable and close to their optimal level since then.

The difference between the HR Medusa1/2 and the IFU/Argus slits is due to different fibre diameters.