DFO Monitors:
CAL
DFO
dfosExplorer
scoreManager
tpManager
External:
mucMonitor
QC external
QC internal
HealthChecks
WISQ
CalibMaps
GIRAFFE HC ScoreManager (report: FFLOC_ALL)
back to:
main
Last update: 2017-10-28T18:00:15 (UT)
GIRAFFE news:
Report news:
If you have edited the report news, they are visible on the Web
but not locally. Then you need to refresh the page.
HC report
edit config.tp file
view trendPlotter config.tp file
1. Score results
This is the score table which displays on the QUICK-LOOK report version; click on score symbols to connect to score report.
FF localization, all settings (scores, last 7 days up to 2017-10-27*)
1:
Med1_all
7
|
2:
Med2_all
9
|
4:
IFU1_all
4
|
5:
IFU2_all
2
|
6:
Argus_all
9
|
Quality comments (certified nights only):
Plot
Date*
AB name (linked to score report)
score
product quality comment
Plot 1
2017-10-25
GIRAF.2017-10-26T13:56:05.464_tpl.ab
SIMCAL fibres saturated because of lamp too bright and slow mode
Plot 1
2017-10-25
GIRAF.2017-10-26T18:14:35.267_tpl.ab
SIMCAL fibres saturated because of lamp too bright and slow mode
Plot 1
2017-10-26
GIRAF.2017-10-27T17:57:33.545_tpl.ab
saturation in the SIMCAL fibres, because of SIMCAL lamp brightness and slow read mode
Plot 2
2017-10-25
GIRAF.2017-10-26T14:41:49.498_tpl.ab
SIMCAL fibres saturated because of lamp too bright and slow mode
Plot 2
2017-10-25
GIRAF.2017-10-26T18:42:52.820_tpl.ab
SIMCAL fibres saturated because of lamp too bright and slow mode
Plot 2
2017-10-27
GIRAF.2017-10-28T05:42:57.789_tpl.ab
SIMLAMP fibres saturated
"Date" links to complete AB product page; "AB" links to score report