[an error occurred while processing this directive]
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+R to enforce refresh of scores and dates]
NACO trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2012-11-14T15:37:15 (UT)
close window history: 2004 2005 2006   2007   2008   2009 2010 2011 2012 2013   HEALTH
  ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEAN 881   VAL770.,1200 38 lamp_flux this | last_yr | all lamp flux in ADU/sec/pixel, for J, S27
2 QC1DB MEAN 977   VAL800.,1200. 41 lamp_flux this | last_yr | all lamp flux in ADU/sec/pixel, for H, S27
3 QC1DB MEAN 847   VAL700.,1200. 45 lamp_flux this | last_yr | all lamp flux in ADU/sec/pixel, for Ks, S27
4 QC1DB MEAN 205   VAL170.,300. 18 lamp_flux this | last_yr | all lamp flux in ADU/sec/pixel, for J, S13
5 QC1DB MEAN 234   VAL200.,320. 26 lamp_flux this | last_yr | all lamp flux in ADU/sec/pixel, for H, S13
6 QC1DB MEAN 207   VAL170.,300. 28 lamp_flux this | last_yr | all lamp flux in ADU/sec/pixel, for Ks, S13
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:naco_lampflat
(QC1 database)
dataset:lamp_flux
mean:881 
fixed thresholds:770.0...1200 
N_data plotted:38
[click on plot for closeup]
Plot 2
data source:naco_lampflat
(QC1 database)
dataset:lamp_flux
mean:977 
fixed thresholds:800.0...1200.0 
N_data plotted:41
[click on plot for closeup]
Plot 3
data source:naco_lampflat
(QC1 database)
dataset:lamp_flux
mean:847 
fixed thresholds:700.0...1200.0 
N_data plotted:45
[click on plot for closeup]
Plot 4
data source:naco_lampflat
(QC1 database)
dataset:lamp_flux
mean:205 
fixed thresholds:170.0...300.0 
N_data plotted:18
[click on plot for closeup]
Plot 5
data source:naco_lampflat
(QC1 database)
dataset:lamp_flux
mean:234 
fixed thresholds:200.0...320.0 
N_data plotted:26
[click on plot for closeup]
Plot 6
data source:naco_lampflat
(QC1 database)
dataset:lamp_flux
mean:207 
fixed thresholds:170.0...300.0 
N_data plotted:28
[click on plot for closeup]
This plot

This plots shows the lamp flux in ADU/sec derived from lamp flat fields. The plot is divided in three broad band filters to monitor beside the lamp property also the throughput per filter.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up