OMEGACAM trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2015-04-15T03:33:57 (UT)
same group: non-linearity   num_NL_pixels linearity_minmax FP_noise
close window history: 2011 2012 2013 2014   2015   2016   2017 2018 2019 2020 2021 2022 2023 2024   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN -0.0147 ADU VAL-0.1,0.1 13 qc_lin_eff this | last_yr | all effective non-linearity correction averaged over all 32 detectors
2 QC1DB MEDIAN -0.0144 ADU none  401 qc_lin_eff this | last_yr | all effective non-linearity correction for each individual detector
3 QC1DB MEDIAN 1e+04 ADU VAL5000.,15000. 13 qc_lin_eff_flux this | last_yr | all flux level at which effective non-linearity correction is applied averaged over all detectors (ADU)
4 LOCAL none   none none  383 det_id n/a score per detector:
GREEN = within the thresholds
4 LOCAL none   none none  0 det_id n/a score per detector:
RED = beyond the thresholds
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
data source:omegacam_detlin
(QC1 database)
dataset:qc_lin_eff
median:-0.0147ADU
fixed thresholds:-0.1...0.1ADU
N_data plotted:13
[click on plot for closeup]
Plot 2
data source:omegacam_detlin
(QC1 database)
dataset:qc_lin_eff
median:-0.0144ADU
thresholds:none 
N_data plotted:401
[click on plot for closeup]
Plot 3
data source:omegacam_detlin
(QC1 database)
dataset:qc_lin_eff_flux
median:1e+04ADU
fixed thresholds:5000.0...15000.0ADU
N_data plotted:13
[click on plot for closeup]
Plot 4
data source:local data source
dataset:
(numbers below apply to this dataset)
det_id
average:none
thresholds:none
N_data plotted:383
[click on plot for closeup]
This plot

Trending plots for the detector linearity.

QC parameters are derived from each of 32 detectors of OmegaCAM dome flat frames.

DB Name: omegacam_detlin
Fits header: QC.LIN.EFF
QC1 DB: qc_lin_eff / qc_num_bpm

Description:
The detector linearity as measured from the detector monitoring pipeline product DET_LIN_INFO and BP_MAP_NL

Setup:
BIN.X / BIN.Y / READ.MODE / READ.SPEED = 1/1/normal/normal

These Health Check plots describe:
Fig. 1 non-linearity_AVG: The effective non-linearity correction averaged over all 32 detectors. This plot is sensitive to any changes common to all detectors.
Fig. 2 non-linearity_ALL: The effective non-linearity correction for each individual detector. This plot is sensitive changes in individual detectors.
Fig. 3 flux_level_AVG: The flux level at which the effective non-linearity correction is applied averaged over all 32 detectors (ADU).
Fig. 4 score_outliers: Scores of the effective non-linearity correction as measured for each of the 32 detectors.
Green = within the defined thresholds.
Red = beyond the defined thresholds.


OmegaCAM detector plane layout (detector_ID locations):
OmegaCAM detector layout

 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up