OMEGACAM trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2021-04-15T03:34:46 (UT)
same group: dark
close window history: 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020   2021   2022   2023 2024   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN -1.07 ADU/pix/hr VAL0.,0.7 14 qc_dark_current this | last_yr | all DARK current averaged over all 32 detectors (ADU/pixel/hour)
2 QC1DB MEDIAN 0.316 ADU/pix/hr none  434 qc_dark_current this | last_yr | all DARK current for each individual detector (ADU/pixel/hour)
3 QC1DB MEDIAN 106 part/cm2/hr VAL100.,120. 14 qc_particle_rate this | last_yr | all particle rate averaged over all 32 detectors (particles/cm2/hour)
4 LOCAL none   none none  420 det_id n/a score per detector:
GREEN = within the thresholds
4 LOCAL none   none none  28 det_id n/a score per detector:
RED = beyond the thresholds
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
data source:omegacam_dark
(QC1 database)
dataset:qc_dark_current
median:-1.07ADU/pix/hr
fixed thresholds:0.0...0.7ADU/pix/hr
N_data plotted:14
[click plot for closeup]
Plot 2
data source:omegacam_dark
(QC1 database)
dataset:qc_dark_current
median:0.316ADU/pix/hr
thresholds:none 
N_data plotted:434
[click plot for closeup]
Plot 3
data source:omegacam_dark
(QC1 database)
dataset:qc_particle_rate
median:106part/cm2/hr
fixed thresholds:100.0...120.0part/cm2/hr
N_data plotted:14
[click plot for closeup]
Plot 4
data source:local data source
dataset:
(numbers below apply to this dataset)
det_id
average:none
thresholds:none
N_data plotted:420
[click plot for closeup]
This plot

Trending plots for the detector dark current and particle rate.

QC parameters are derived from each of 32 detectors of OmegaCAM dark frames.

DB Name: omegacam_dark
Fits header: QC.DARK.CURRENT / QC.PARTICLE.RATE
QC1 DB: qc_dark_current / qc_particle_rate

Description:
The dark current and particle rate is measured from the pipeline product MEAN_DARK

Setup:
BIN.X / BIN.Y / READ.MODE / READ.SPEED = 1/1/normal/normal

These Health Check plots describe:
Fig. 1 dark_current_AVG: The dark current averaged over all 32 detectors (in ADU/pixel/hour). This plot is sensitive to any changes common to all detectors.
Fig. 2 gain_ALL: The dark current measured for each individual detector (in ADU/pixel/hour). This plot is sensitive changes in individual detectors.
Fig. 3 particle_rate_AVG: The dark frame particle rate averaged over all 32 detectors (in particles/cm^2/hour).
Fig. 4 score_outliers: Scores of the dark current as measured for each of the 32 detectors.
Green = within the defined thresholds.
Red = beyond the defined thresholds.


OmegaCAM detector plane layout (detector_ID locations):
OmegaCAM detector layout

 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up