SINFONI trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2013-01-18T10:14:45 (UT)
same group: DARK_dit=300   dit=10 dit=60 dit=600 dit=900
close window history: 2005 2006 2007 2008 2009 2010   2011   2012   2013 2014 2015 2016 2017 2018 2019   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 10.5 ADU VAL0.0,25.0 65 qc_darkmed_ave this | last_yr | all median dark level, as computed from the master dark
2 QC1DB MEDIAN 0.682 ADU VAL-3.0,4.0 65 qc_darkmed_stdev this | last_yr | all standard deviation of median dark level
3 QC1DB MEDIAN 3.01 ADU VAL2.5,3.75 65 qc_ron2 this | last_yr | all rms noise derived by subtracting 2 consecutive raw darks (normalized to DET.DIT = 1 second)
4 QC1DB MEDIAN 2.42 ADU VAL1.0,4.0 65 qc_gaf_fpn this | last_yr | all fixed pattern noise as computed by fitting a Gaussian to the master dark histogram
5 QC1DB MEDIAN 1.02e+04 number VAL2000,20000 65 qc_bpm_nbadpix this | last_yr | all number of pixels with a high dark current
6 QC1DB MEDIAN 0.00892 unitless VAL-0.2,0.3 65 qc_oes1 this | last_yr | all  
7 QC1DB MEDIAN -0.0252 unitless VAL-0.2,0.3 65 qc_oes2 this | last_yr | all  
8 QC1DB MEDIAN -0.0105 unitless VAL-0.2,0.3 65 qc_oes3 this | last_yr | all  
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:sinfoni_dark
(QC1 database)
dataset:qc_darkmed_ave
median:10.5ADU
fixed thresholds:0.0...25.0ADU
N_data plotted:65
[click on plot for closeup]
Plot 2
data source:sinfoni_dark
(QC1 database)
dataset:qc_darkmed_stdev
median:0.682ADU
fixed thresholds:-3.0...4.0ADU
N_data plotted:65
[click on plot for closeup]
Plot 3
data source:sinfoni_dark
(QC1 database)
dataset:qc_ron2
median:3.01ADU
fixed thresholds:2.5...3.75ADU
N_data plotted:65
[click on plot for closeup]
Plot 4
data source:sinfoni_dark
(QC1 database)
dataset:qc_gaf_fpn
median:2.42ADU
fixed thresholds:1.0...4.0ADU
N_data plotted:65
[click on plot for closeup]
Plot 5
data source:sinfoni_dark
(QC1 database)
dataset:qc_bpm_nbadpix
median:1.02e+04number
fixed thresholds:2000...20000number
N_data plotted:65
[click on plot for closeup]
Plot 6
data source:sinfoni_dark
(QC1 database)
dataset:qc_oes1
median:0.00892unitless
fixed thresholds:-0.2...0.3unitless
N_data plotted:65
[click on plot for closeup]
Plot 7
data source:sinfoni_dark
(QC1 database)
dataset:qc_oes2
median:-0.0252unitless
fixed thresholds:-0.2...0.3unitless
N_data plotted:65
[click on plot for closeup]
Plot 8
data source:sinfoni_dark
(QC1 database)
dataset:qc_oes3
median:-0.0105unitless
fixed thresholds:-0.2...0.3unitless
N_data plotted:65
[click on plot for closeup]
This plot

These are the trending plots for the SINFONI QC parameters derived from DARK frames.

These plots describe:

1. the median DARK level, as computed from each master dark frame.

2. the standard deviation of the median DARK level.

3. the root-mean-squared noise, as derived by subtracting two consecutive raw dark frames, computing a median standard deviation in
selected sub-windows, computing a median standard deviation in selected sub-windows in the difference image, and then normalizing to DET.DIT=1 second.

4. the fixed pattern noise, as computed by fitting a Gaussian to the histogram of the master dark frames. The fpn is simply the standard deviation of Gaussian.

5. the number of pixels with a high dark current.

6/7/8. the odd-even column stripes in the area of slitlet #25. If med(odd) = median of all odd columns in the area of slitlet #25
and med(even) = median of all even columns in the area of slitlet #25 then, this plot is defined as:
oes25 = 2 * (med(odd) - med(even)) / (med(odd) + med(even))
the plots show the effect for the first 3 raw files of the DARK

DARK current and RON higher than usual could be due to an increase of the detector temperature (INS.TEMP6.VAL), grating temperature (INS.TEMP13.VAL) or of the rack temperature (INS.TEMP6.VAL). Check also INS.SENS13.ST (Detector warm mode).

Find more information about SINFONI DARK frames here.

Other QC parameters related to detector properties (like dark level, linearity, gain, etc.) are also trended here.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up