[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SINFONI trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2016-07-15T03:30:37 (UT)
same group: lamp_eff   lamp_badpix lamp_off lamp_fpn1 lamp_fpn2 vignetting_check_sl1 vignetting_check_sl32
close window history: 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015   2016   2017   2018 2019   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 oQC1DB none   ADU VAL4000,11750 61 qc_specflux_cavg this | last_yr | all median lamp counts in the J-band at all camera scales
(this FULL data set is scored)
1 QC1DB none   ADU none  45 qc_specflux_cavg this | last_yr | all median lamp counts: J-band/250 mas scale
1 QC1DB none   ADU none  10 qc_specflux_cavg this | last_yr | all median lamp counts: J-band/100 mas scale
1 QC1DB none   ADU none  6 qc_specflux_cavg this | last_yr | all median lamp counts: J-band/25 mas scale

2 oQC1DB none   ADU VAL4000,11750 42 qc_specflux_cavg this | last_yr | all median lamp counts in the H-band at all camera scales
(this FULL data set is scored)
2 QC1DB none   ADU none  24 qc_specflux_cavg this | last_yr | all median lamp counts in the H-band/250 mas scale
2 QC1DB none   ADU none  14 qc_specflux_cavg this | last_yr | all median lamp counts in the H-band/100 mas scale
2 QC1DB none   ADU none  4 qc_specflux_cavg this | last_yr | all median lamp counts in the H-band/25 mas scale

3 oQC1DB none   ADU VAL4000,11750 88 qc_specflux_cavg this | last_yr | all median lamp counts in the K-band at all camera scales
(this FULL data set is scored)
3 QC1DB none   ADU none  31 qc_specflux_cavg this | last_yr | all median lamp counts in the K-band/250 mas scale
3 QC1DB none   ADU none  28 qc_specflux_cavg this | last_yr | all median lamp counts in the K-band/100 mas scale
3 QC1DB none   ADU none  29 qc_specflux_cavg this | last_yr | all median lamp counts in the K-band/25 mas scale

4 oQC1DB none   ADU VAL4000,11500 65 qc_specflux_cavg this | last_yr | all median lamp counts in the H+K-band at all camera scales
(this FULL data set is scored)
4 QC1DB none   ADU none  43 qc_specflux_cavg this | last_yr | all median lamp counts in the H+K-band/250 mas scale
4 QC1DB none   ADU none  13 qc_specflux_cavg this | last_yr | all median lamp counts in the H+K-band/100 mas scale
4 QC1DB none   ADU none  9 qc_specflux_cavg this | last_yr | all median lamp counts in the H+K-band/25 mas scale
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_specflux_cavgo
average:none
fixed thresholds:4000...11750ADU
N_data plotted:61
[click on plot for closeup]
Plot 2
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_specflux_cavgo
average:none
fixed thresholds:4000...11750ADU
N_data plotted:42
[click on plot for closeup]
Plot 3
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_specflux_cavgo
average:none
fixed thresholds:4000...11750ADU
N_data plotted:88
[click on plot for closeup]
Plot 4
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_specflux_cavgo
average:none
fixed thresholds:4000...11500ADU
N_data plotted:65
[click on plot for closeup]
This plot

These are the trending plots for the SINFONI QC parameters derived from the LAMP FLATs.

These plots describe:
Day-time FLAT calibrations come in at least five pairs of lamp on/off frames. For each
raw lamp-on frame the corresponding lamp-off frame is subtracted and the median is calculated.
The above lamp efficiencies are the average of these 5 (or more) values.

Median lamp count limits should remain between 4,000 and 10,000 ADU. The lower limit ensures that the flux levels of the flats are sufficiently high for good statistical noise limits, while the upper limit of 10,000 counts should not be exceeded as this can approach the detector's non-linear regime.

1. the median lamp counts in the J-band for the camera scales: 250, 100, and 25 mas
2. the median lamp counts in the H-band for the camera scales: 250, 100, and 25 mas
3. the median lamp counts in the K-band for the camera scales: 250, 100, and 25 mas
4. the median lamp counts in the H+K-band for the camera scales: 250, 100, and 25 mas

Find more information about SINFONI FLAT frames here.

 


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )