[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SINFONI trending system: FULL report
HELP USERS-GUIDE MORE
Last update: 2014-09-01T14:34:22 (UT)
  | now: 2014-09-01T15:02:23 (UT) muc04 QC pipeline: sinfo-2.5.4 (installed 2014-08-22)
close window history: 2005 2006 2007 2008 2009 2010 2011 2012 2013   2014   HEALTH FULL
  ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (3) none   ADU VAL4000,10000 671 qc_specflux_cavg all (full time range) median lamp counts in the J-band at the 250 mas camera scale
1 oQC1DB yes (3) none   ADU VAL4000,10000 128 qc_specflux_cavg all (full time range) median lamp counts in the J-band at the 100 mas camera scale
1 QC1DB yes (3) none   ADU VAL4000,10000 82 qc_specflux_cavg all (full time range) median lamp counts in the J-band at the 25 mas camera scale
2 QC1DB yes (4) none   ADU VAL4000,10000 628 qc_specflux_cavg all (full time range) median lamp counts in the H-band at the 250 mas camera scale
2 oQC1DB yes (4) none   ADU VAL4000,10000 250 qc_specflux_cavg all (full time range) median lamp counts in the H-band at the 100 mas camera scale
2 QC1DB yes (4) none   ADU VAL4000,10000 48 qc_specflux_cavg all (full time range) median lamp counts in the H-band at the 25 mas camera scale
3 QC1DB yes (7) none   ADU VAL4000,10000 912 qc_specflux_cavg all (full time range) median lamp counts in the K-band at the 250 mas camera scale
3 oQC1DB yes (7) none   ADU VAL4000,10000 851 qc_specflux_cavg all (full time range) median lamp counts in the K-band at the 100 mas camera scale
3 QC1DB yes (7) none   ADU VAL4000,10000 425 qc_specflux_cavg all (full time range) median lamp counts in the K-band at the 25 mas camera scale
4 QC1DB yes (7) none   ADU VAL4000,10000 779 qc_specflux_cavg all (full time range) median lamp counts in the H+K-band at the 250 mas camera scale
4 oQC1DB yes (7) none   ADU VAL4000,10000 385 qc_specflux_cavg all (full time range) median lamp counts in the H+K-band at the 100 mas camera scale
4 QC1DB yes (7) none   ADU VAL4000,10000 363 qc_specflux_cavg all (full time range) median lamp counts in the H+K-band at the 25 mas camera scale
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: not implemented
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_specflux_cavg
average:none
fixed thresholds:4000...10000ADU
N_data plotted:671
[click on plot for closeup]
Plot 2
scores: not implemented
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_specflux_cavg
average:none
fixed thresholds:4000...10000ADU
N_data plotted:628
[click on plot for closeup]
Plot 3
scores: not implemented
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_specflux_cavg
average:none
fixed thresholds:4000...10000ADU
N_data plotted:912
[click on plot for closeup]
Plot 4
scores: not implemented
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_specflux_cavg
average:none
fixed thresholds:4000...10000ADU
N_data plotted:779
[click on plot for closeup]
This plot

These are the trending plots for the SINFONI QC parameters derived from the LAMP FLATs.

These plots describe:
Day-time FLAT calibrations come in at least five pairs of lamp on/off frames. For each
raw lamp-on frame the corresponding lamp-off frame is subtracted and the median is calculated.
The above lamp efficiencies are the average of these 5 (or more) values.

Median lamp count limits should remain between 4,000 and 10,000 ADU. The lower limit ensures that the flux levels of the flats are sufficiently high for good statistical noise limits, while the upper limit of 10,000 counts should not be exceeded as this can approach the detector's non-linear regime.

1. the median lamp counts in the J-band for the camera scales: 250, 100, and 25 mas
2. the median lamp counts in the H-band for the camera scales: 250, 100, and 25 mas
3. the median lamp counts in the K-band for the camera scales: 250, 100, and 25 mas
4. the median lamp counts in the H+K-band for the camera scales: 250, 100, and 25 mas

Find more information about SINFONI FLAT frames here.

 


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPS column.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up