[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+R to enforce refresh of scores and dates]
SINFONI trending system: FULL report
Last update: 2012-05-20T15:01:04 (UT)
| now: 2012-05-24T13:29:11 (UT)
HELP USERS-GUIDE MORE
same group: detector_gain   nonlinear_pixels linearity_coef_C0 coef_C1 coef_C2
General news: 2012-05-24: The HC monitors for OMEGACAM, VIRCAM and HAWKI are back. Population with current data is in progress.
SINFONI news:
Report news:  
DATE*: [?] 2012-05-17
report | NR
2012-05-18
report | NR
2012-05-19
report | NR
2012-05-20
report | NR
2012-05-21
report | NR
2012-05-22
report | NR
2012-05-23
report | NR
P...
Product quality:
products products products products products products products
  no OPSLOG data  
close window history: 2005 2006 2007 2008 2009 2010   2011   HEALTH FULL
  ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB no MEDIAN 2.32977 e-/ADU VAL1.75,2.95 224 qc_gain all (full time range) detector gain in the J-band (only for the 25 mas camera scale)
2 QC1DB no MEDIAN 2.36125 e-/ADU VAL1.75,2.95 210 qc_gain all (full time range) detector gain in the H-band (only for the 25 mas camera scale)
3 QC1DB no MEDIAN 2.40442 e-/ADU VAL1.75,2.95 324 qc_gain all (full time range) detector gain in K-band (only for the 25 mas camera scale)
4 QC1DB no MEDIAN 2.33437 e-/ADU VAL1.75,2.95 296 qc_gain all (full time range) detector gain in the H+K band (only for the 25 mas camera scale)
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
Plot 1
scores: not implemented
data source:sinfoni_line
(QC1 database)
dataset:qc_gain
median:2.32977e-/ADU
fixed thresholds:1.75...2.95e-/ADU
N_data plotted:224
[click on plot for closeup]
Plot 2
scores: not implemented
data source:sinfoni_line
(QC1 database)
dataset:qc_gain
median:2.36125e-/ADU
fixed thresholds:1.75...2.95e-/ADU
N_data plotted:210
[click on plot for closeup]
Plot 3
scores: not implemented
data source:sinfoni_line
(QC1 database)
dataset:qc_gain
median:2.40442e-/ADU
fixed thresholds:1.75...2.95e-/ADU
N_data plotted:324
[click on plot for closeup]
Plot 4
scores: not implemented
data source:sinfoni_line
(QC1 database)
dataset:qc_gain
median:2.33437e-/ADU
fixed thresholds:1.75...2.95e-/ADU
N_data plotted:296
[click on plot for closeup]
General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up