SINFONI trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2014-12-19T02:04:08 (UT)
  | now: 2014-12-19T04:05:58 (UT) muc04 QC pipeline: sinfo-2.5.4 (installed 2014-10-20)
same group: slit_dist_ALL slit_dist_JHK distortion_shift   distortion_coef00 coef01 coef10 coef11 coef20 coef02 coef21
close window history: 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014   HEALTH
  ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB no MEDIAN 14.3 pixels VAL10.0,18.0 51 qc_coeff_00 this | last_yr | all J-band distortion coefficient C00 (dist = C00 + C10x + C01y + C11xy + C20x^2 + C02y^2 + C21x^2y + C12xy^2)
2 QC1DB no MEDIAN 17.9 pixels VAL14.0,22.0 35 qc_coeff_00 this | last_yr | all H-band distortion coefficient C00
3 QC1DB no MEDIAN 23.9 pixels VAL20.0,28.0 59 qc_coeff_00 this | last_yr | all K-band distortion coefficient C00
4 QC1DB no MEDIAN 21.3 pixels VAL17.0,25.0 92 qc_coeff_00 this | last_yr | all H+K-band distortion coefficient C00
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: 2|
data source:sinfoni_dist
(QC1 database)
dataset:qc_coeff_00
median:14.3pixels
fixed thresholds:10.0...18.0pixels
N_data plotted:51
[click on plot for closeup]
Plot 2
scores: 1|
data source:sinfoni_dist
(QC1 database)
dataset:qc_coeff_00
median:17.9pixels
fixed thresholds:14.0...22.0pixels
N_data plotted:35
[click on plot for closeup]
Plot 3
scores: 2|
data source:sinfoni_dist
(QC1 database)
dataset:qc_coeff_00
median:23.9pixels
fixed thresholds:20.0...28.0pixels
N_data plotted:59
[click on plot for closeup]
Plot 4
scores: 2|
data source:sinfoni_dist
(QC1 database)
dataset:qc_coeff_00
median:21.3pixels
fixed thresholds:17.0...25.0pixels
N_data plotted:92
[click on plot for closeup]
This plot

These are the trending plots for the SINFONI QC parameters derived from DISTORTION frames.

Distortion frames are obtained for the J, H, K, and H+K bandpasses, but only for the 250 mas camera.

 


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up