UVES Quality Control:
Dark

QUALITY CONTROL
HOME
UVES QC
Trending & QC1
   Dark
Pipeline
Data Packages
Data Management
QC links:
uves_dark :: Calculated
 
QC PLOTS
  CURRENT HISTORY
Daily Health Checks:    
DARK SM
QC1 uves_dark database (advanced users): browse | plot
   Click on CURRENT to see the last trending (Health Check).
   Click on HISTORY to see the historical evolution of the trending.


top Calculated
The following QC parameters are currently calculated by the QC report scripts, not by the pipeline. Eventually the pipeline should calculate these quantities.

QC1 parameters

  • mean_dark normalized to 3600 secs exposure time (QC1 database table uves_dark, column current_dark

 

DARK frames

Modes

DARK frames (CATG = CALIB, TYPE = DARK, TECH = IMAGE) are measured occasionally, for the BLUE and RED arms separately, with the shutter closed. They are used to measure the dark current. They come in 1x1 and 2x2 binnings. They have typical exposure times of 1 hr.

There are also DARKs with short exposure times, typically 60 secs. These are used for CCD cleanup and are useless for monitoring the dark current.

There are also open-shutter DARKs (since December 2001: CATG = CALIB, TYPE = PARASITIC, TECH = IMAGE). They include, in addition to the CCD dark current, contribution from the camera enclosure. During commissioning, there has been found a level of luminiscence in the blue camera which, on data from March 2001, is still persistent. It amounts to roughly 11 ADU/hr/4px or about 1.7 e-/hr/px. Hence this contribution dominates the effective dark current in the blue. It most likely originates from Cerenkov radiation due to radioactive decay.


QC1 parameters

Dark current. The UVES pipeline generates master DARKs from input raw DARKs. A master BIAS is subtracted. The mean of the residual pixels is calculated and transformed into the dark current.

Usually results from the 1x1 darks are unreliable since the dark current is actually very low. Hence all numbers given here are for the 2x2 bin mode. They are in units of [ADU/hr/4px]. The transformed values in electrons per hour have used the conversion factors 0.60 electrons per ADU for the blue, and 0.52 electrons per ADU for the red CCDs.

parameter procedure applicable typical values * units plot**
CCD dark current mean across whole CCD master, raw
BLUE_2x2:
 
2.1
REDU_2x2:
 
2.9
REDL_2x2:
 
2.4
ADU/hr/4px 1
     
BLUE
 
0.3
REDU
 
0.4
REDL
 
0.3
e-/hr/px  
CCD dark current (shutter open) same same
BLUE_2x2:
 
13.4
REDU_2x2:
 
3.9
REDL_2x2:
 
4.3
ADU/hr/4px 2
     
BLUE
 
2.0
REDU
 
0.5
REDL
 
0.6
e-/hr/px  
parasitic light values from '2' minus values from '1'  
BLUE
 
1.7
REDU
 
0.1:
REDL
 
0.2:
e-/hr/px (2-1)

* mean for 2001
** refers to numbering in QC plot

 


 