uves_fib_wave :: Fibre-Resolution-Test-Results
 
QC PLOTS
  CURRENT HISTORY
Daily Health Checks:    
MOS WAVE ResolvingPower 580 7+1 U DHC
MOS WAVE ResolvingPower 580 7+1 L DHC
MOS WAVE Dispersion 580 7+1 U DHC
MOS WAVE Dispersion 580 7+1 L DHC
MOS WAVE NumLines 580 7+1 U DHC
MOS WAVE NumLines 580 7+1 L DHC
MOS WAVE WLmin 580 7+1 U DHC
MOS WAVE WLmin 580 7+1 L DHC
MOS WAVE WLmax 580 7+1 U DHC
MOS WAVE WLmax 580 7+1 L DHC
QC1 uves_fib_wave database (advanced users): browse | plot
QC1 uves_fib_wave_scoreQC database (advanced users): browse | plot
   Click on CURRENT to see the last trending (Health Check).
   Click on HISTORY to see the historical evolution of the trending.


top Introduction

This tutorial provides information for the MOS - Wavelength Solution Health Check trending.
The MOS - Wavelength Solution trending is based on the measurement of QC1 parameters of MOS WAVECAL calibration data obtained with a dedicated Health Check OB executed daily. This OBs obtains one set of MOS WAVECAL data in the 7+1 Fibre mode, 580.0nm wavelength MOS setting useing Plate-1.

top Fibre-Resolution-Test-Results

QC1 parameters

  • median resolving power of lines selected (QC1 database table uves_fib_wave, column resol_med, product keyword QC.RESOLMED)
    Description: measured median resolving power of lines selected
  • sigma of residual of line positions to fit [10**-12 m] (QC1 database table uves_fib_wave, column resid_rms, product keyword QC.LINE.RESIDRMS)
    Description: Parameter useful to check precision of wavelength calibration solution We use picometer units at it comes out a very small number.
  • number of lines selected (QC1 database table uves_fib_wave, column nlin_sel, product keyword QC.NLINSEL)
    Description: number of lines which could be cross-identified with the catalogue

 

 


 