![]() |
|||||
|---|---|---|---|---|---|
| UVES@KUEYEN:
QC1 values from WAVE frames |
|||||
|
|
|||||||||||||||||||||||||||||||||
| WAVE frames | |||||||||||||||||||||||||||||||||
|
Modes Parameters are measured for all three CCDs separately (BLUE, REDL, REDU). They are available whenever raw arclamp (type LAMP,WAVE) frames are measured in a UVES standard mode. Since Period 68 (Oct. 01, 2001) a set of daily arclamp exposures is taken which serves as instrument health check, in the two settings BLUE 437 1x1 and RED 580 1x1 and with slit widths 0.4 and 0.3 arcsec, resp. These are used to monitor the resolving power of UVES. QC1 parameters: resolving power, dispersion, line intensity The most important QC1 parameters measured are resolving power R and standard deviation of the dispersion solution, sigma. Others are e.g. minimum and maximum wavelength and number of lines selected. Resolving power R = lambda/delta lambda is evaluated as median over a Gauss fit to all selected lines. R is compared to reference values as function of slit width. The reference values are lab data as taken from the User's Manual. The quality of the dispersion solution is measured by the standard deviation of the difference between the found and fitted position, again evaluated for all selected lines. Typically 300 - 1000 lines are selected per CCD.
|
|||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||
|
* in Oct-Dec 2000 FORMAT All standard settings have one box, separated for 1x1 and 2x2 binnings. Resolving power values are collected in resolution plot 1 regardless of the associated slit width.
|
|||||||||||||||||||||||||||||||||
|
|