[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
UVES trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2021-07-15T07:00:29 (UT)
same group: Median Master_RON Raw_RON Struct_X Struct_Y Pattern_noise
close window history: 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020   2021   2022   2023 2024   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 193 ADU none  104 median_master this | last_yr | all median level in MASTER BIAS, readout mode 225kHz,1x1,low; BLUE CCD
2 QC1DB MEDIAN 206 ADU none  145 median_master this | last_yr | all same, for REDL CCD
3 QC1DB MEDIAN 233 ADU none  145 median_master this | last_yr | all same, for REDU CCD
4 QC1DB MEDIAN 193 ADU none  106 median_master this | last_yr | all median level in MASTER BIAS, readout mode 50kHz,2x2,high; BLUE CCD
5 QC1DB MEDIAN 207 ADU none  105 median_master this | last_yr | all same, for REDL CCD
6 QC1DB MEDIAN 278 ADU none  105 median_master this | last_yr | all same, for REDU CCD
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
data source:uves_bias
(QC1 database)
dataset:median_master
median:193ADU
thresholds:none 
N_data plotted:104
[click plot for closeup]
Plot 2
data source:uves_bias
(QC1 database)
dataset:median_master
median:206ADU
thresholds:none 
N_data plotted:145
[click plot for closeup]
Plot 3
data source:uves_bias
(QC1 database)
dataset:median_master
median:233ADU
thresholds:none 
N_data plotted:145
[click plot for closeup]
Plot 4
data source:uves_bias
(QC1 database)
dataset:median_master
median:193ADU
thresholds:none 
N_data plotted:106
[click plot for closeup]
Plot 5
data source:uves_bias
(QC1 database)
dataset:median_master
median:207ADU
thresholds:none 
N_data plotted:105
[click plot for closeup]
Plot 6
data source:uves_bias
(QC1 database)
dataset:median_master
median:278ADU
thresholds:none 
N_data plotted:105
[click plot for closeup]
This plot

This is the trending report for the BIAS level in UVES frames. Note that this parameter is based on the keyword PRO.DATAMED which is not present in the QC1 opslog files.
These plots describe:

  1. the BIAS level in masterbias frames of the BLUE CCD in the 1x1 binning, 225kHz, low gain readout mode
  2. same for RED LOWER CCD
  3. same for RED UPPER CCD
  4. the BIAS level in masterbias frames of the BLUE CCD in the 2x2 binning, 50kHz, high gain readout mode
  5. same for RED LOWER CCD
  6. same for RED UPPER CCD

The thresholds are set using the KPI document : "UVES critical Quality parameters"
The higher thrshold is always set to 325. The lower threshold depends of the Mode:
  1. B 1x1 : 27
  2. B 2x2 : 49
  3. U 1x1 : 33
  4. U 2x2 : 60
  5. L 1x1 : 36
  6. L 2x2 : 62
The reference values are the following
  1. B 1x1 : 195
  2. B 2x2 : 196
  3. U 1x1 : 204
  4. U 2x2 : 205
  5. L 1x1 : 202
  6. L 2x2 : 206
Find more information about UVES BIAS frames here.
Other QC parameters related to detector properties (like linearity, gain) are trended here.
 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up