UVES trending system: FULL report
HELP USERS-GUIDE MORE
Last update: 2024-04-25T02:05:31 (UT)
  | now: 2024-04-25T05:40:53 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
close window history: 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
- for reprocessing data: QC1_browser | QC1_plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB none   counts/sec none  1318 mean_master all (full time range) Stability of LAMP1 (flatfield deuterium lamp), BLUE CCD, 346nm, 1x1
1 QC1DB none   counts/sec none  216 mean_master all (full time range) same, from reprocessing 2015 with uves/5.5.3
2 QC1DB none   counts/sec none  1759 mean_master all (full time range) same for LAMP3 (FF lamp), same setups
2 QC1DB none   counts/sec none  350 mean_master all (full time range) same, from reprocessing 2015 with uves/5.5.3
3 QC1DB none   counts/sec none  1833 mean_master all (full time range) same for LAMP4 (FF lamp), 437nm, 1x1

3 QC1DB none   counts/sec none  617 mean_master all (full time range) same, from reprocessing 2015 with uves/5.5.3
5 QC1DB none   counts/sec none  1725 mean_master all (full time range) same for LAMP5 (FF lamp), RED UPPER CCD, 580nm, 1x1
5 QC1DB none   counts/sec none  581 mean_master all (full time range) same, from reprocessing 2015 with uves/5.5.3
6 QC1DB none   counts/sec none  1725 mean_master all (full time range) same for LAMP5, RED LOWER CCD, 580nm, 1x1
6 QC1DB none   counts/sec none  576 mean_master all (full time range) same, from reprocessing 2015 with uves/5.5.3
7 QC1DB none   counts/sec none  1242 mean_master all (full time range) same for LAMP6 (FF lamp), RED UPPER CCD, 860nm, 1x1
7 QC1DB none   counts/sec none  297 mean_master all (full time range) same, from reprocessing 2015 with uves/5.5.3
8 QC1DB none   counts/sec none  1242 mean_master all (full time range) same for LAMP6, RED LOWER CCD, 860nm, 1x1
8 QC1DB none   counts/sec none  298 mean_master all (full time range) same, from reprocessing 2015 with uves/5.5.3
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:uves_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
mean_master
average:none
thresholds:none
N_data plotted:1318
[click plot for closeup]
Plot 2
scores: not implemented
data source:uves_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
mean_master
average:none
thresholds:none
N_data plotted:1759
[click plot for closeup]
Plot 3
scores: not implemented
data source:uves_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
mean_master
average:none
thresholds:none
N_data plotted:1833
[click plot for closeup]
Plot 5
scores: not implemented
data source:uves_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
mean_master
average:none
thresholds:none
N_data plotted:1725
[click plot for closeup]
Plot 6
scores: not implemented
data source:uves_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
mean_master
average:none
thresholds:none
N_data plotted:1725
[click plot for closeup]
Plot 7
scores: not implemented
data source:uves_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
mean_master
average:none
thresholds:none
N_data plotted:1242
[click plot for closeup]
Plot 8
scores: not implemented
data source:uves_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
mean_master
average:none
thresholds:none
N_data plotted:1242
[click plot for closeup]
This plot

This is the trending report for the UVES flat lamps. It monitors the deuterium lamp and the four standard flat-field lamps. It is based on data taken roughly once a week. Trends in this plot probably indicate aging of the lamp.
These plots describe:

  1. the mean_master QC parameter for LAMP 1, the Deuterium lamp used for BLUE FLATs with settings CD#1, 300-360nm
  2. the mean_master QC parameter for LAMP 3, used for BLUE FLATs with settings CD#1, 300-400nm
  3. the mean_master QC parameter for LAMP 4, used for BLUE FLATs with settings CD#2, 330-480nm
  4. the mean_master QC parameter for LAMP 5, REDU CCD, used for RED FLATs with settings CD#3, 450-650nm (which also includes MOS slit flats for the 520 and 580 settings)
  5. the same for the REDL CCD
  6. the mean_master QC parameter for LAMP 6, REDU CCD, used for RED FLATs with settings CD#3, 650-750nm and CD#4, 600-950nm (which also includes MOS slit flats for the 860 setting)
  7. the same for the REDL CCD.

Data points in blue refer to the master calibration reprocessing project UVES_M from 2015.
Find more information about UVES Echelle FLAT frames here.  
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up