UVES trending system: HEALTH CHECK report
Last update: 2012-02-10T09:45:17 (UT)
| now: 2012-02-10T22:08:49 (UT)
HELP USERS-GUIDE MORE
same group: Mean
General news:
UVES news: 2012-02-02 HC plots adjusted to new Y position of red grating
Report news:  
DATE*: [?] 2012-02-03
report | NR
2012-02-04
report | NR
2012-02-05
report | NR
2012-02-06
report | NR
2012-02-07
report | NR
2012-02-08
report | NR
2012-02-09
report | NR
2012-02-10
report | NR
P...
Product quality:
products products products products products products products products
  last OPSLOG data:
2012-02-09*

last files ...
monitor

scores&comments | history ... | plot tutorial ... | contact                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (11) FIXED 0.0 Pix OFF± 0.6 86 ydiffavg this | last_yr | all Mean of Y-Differences between the two frames, CD#1
2 QC1DB yes (11) FIXED 0.0 Pix OFF± 0.6 86 ydiffavg this | last_yr | all same, for CD#2
3 QC1DB yes (11) FIXED 0.0 Pix OFF± 0.4 86 ydiffavg this | last_yr | all same, for CD#3
4 QC1DB yes (11) FIXED 0.0 Pix OFF± 0.4 86 ydiffavg this | last_yr | all same, for CD#4
5 QC1DB yes (11) MEAN 0.00861 Pix VAL0.005,0.015 86 ydiffrms this | last_yr | all RMS of Y-Differences between the two frames
6 QC1DB yes (11) MEAN 0.02392 Pix VAL0.010,0.040 86 ydiffrms this | last_yr | all same, for CD#2
7 QC1DB yes (11) MEAN 0.0191 Pix VAL0.010,0.030 86 ydiffrms this | last_yr | all same, for CD#3
8 QC1DB yes (11) MEAN 0.03652 Pix VAL0.025,0.055 86 ydiffrms this | last_yr | all same, for CD#4
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
Plot 1
scores: no current data
data source:uves_zflat
(QC1 database)
dataset:ydiffavg
fixed average:0.0Pix
rel. thresholds: -0.6...0.6Pix
N_data plotted:86
[click on plot for closeup]
Plot 2
scores: no current data
data source:uves_zflat
(QC1 database)
dataset:ydiffavg
fixed average:0.0Pix
rel. thresholds: -0.6...0.6Pix
N_data plotted:86
[click on plot for closeup]
Plot 3
scores: 8|
data source:uves_zflat
(QC1 database)
dataset:ydiffavg
fixed average:0.0Pix
rel. thresholds: -0.4...0.4Pix
N_data plotted:86
[click on plot for closeup]
Plot 4
scores: 8|
data source:uves_zflat
(QC1 database)
dataset:ydiffavg
fixed average:0.0Pix
rel. thresholds: -0.4...0.4Pix
N_data plotted:86
[click on plot for closeup]
Plot 5
scores: no current data
data source:uves_zflat
(QC1 database)
dataset:ydiffrms
mean:0.00861Pix
fixed thresholds:0.005...0.015Pix
N_data plotted:86
[click on plot for closeup]
Plot 6
scores: no current data
data source:uves_zflat
(QC1 database)
dataset:ydiffrms
mean:0.02392Pix
fixed thresholds:0.01...0.04Pix
N_data plotted:86
[click on plot for closeup]
Plot 7
scores: 8|
data source:uves_zflat
(QC1 database)
dataset:ydiffrms
mean:0.0191Pix
fixed thresholds:0.01...0.03Pix
N_data plotted:86
[click on plot for closeup]
Plot 8
scores: 8|
data source:uves_zflat
(QC1 database)
dataset:ydiffrms
mean:0.03652Pix
fixed thresholds:0.025...0.055Pix
N_data plotted:86
[click on plot for closeup]
This plot

This is the trending report for UVES QC parameters derived from the Cross-Disperser Alignment reproducibility frames.

These data are taken daily. Each Cross-Disperser [CD] is moved into position such that the zeroth order from the echelle falls on the detector and the first frame is exposed, then the CD is moved away, and then moved back into position and the second frame is taken. Ideally, if the mechanical mechanisms are working perfectly, the zeroth order will appear at exactly the same position in the two frames. So this data tests the reproducability of the instrumental setup, in particular that the instrumental setups used during the night can be accurately reproduced during daytime calibrations.

The spatial position of the zeroth order is then measured at a number (~25) positions along the order in the two frames. Plots 1 to 4 trend the mean of the differences between the measured positions in the two frames. Plots 5-8 trend the RMS of these differences.

Find more information about UVES CD_ALIGN frames here.  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPS column.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up