[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-04-30T19:14:47 (UT)
  | now: 2024-05-01T03:57:32 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: Res_U Res_L   Disp_U Disp_L   NLines_U NLines_L   WLmin_U WLmin_L   WLmax_U WLmax_L
close window history: 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023   2024   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 2.99 none none  80 resid_rms this | last_yr | all Dispersion of wavelength solution, from MOS WAVE data, 580.0nm, 1x1, Plate 1, 7+1FIB Mode, Fibre 1, RED UPPER CCD
2 QC1DB MEDIAN 1.98 none none  81 resid_rms this | last_yr | all same, for fibre 2
3 QC1DB MEDIAN 1.97 none none  81 resid_rms this | last_yr | all same, for fibre 3
4 QC1DB MEDIAN 1.98 none VAL1.0,3.1 81 resid_rms this | last_yr | all same, for fibre 4This plot is not scored;

5 QC1DB MEDIAN 2.03 none VAL1.0,3.1 81 resid_rms this | last_yr | all same, for fibre 5This plot is scored;
6 QC1DB MEDIAN 1.99 none none  81 resid_rms this | last_yr | all same, for fibre 6
7 QC1DB MEDIAN 1.96 none none  81 resid_rms this | last_yr | all same, for fibre 7
8 QC1DB MEDIAN 2.02 none none  81 resid_rms this | last_yr | all same, for fibre 8
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resid_rms
median:2.99none
thresholds:none 
N_data plotted:80
[click plot for closeup]
Plot 2
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resid_rms
median:1.98none
thresholds:none 
N_data plotted:81
[click plot for closeup]
Plot 3
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resid_rms
median:1.97none
thresholds:none 
N_data plotted:81
[click plot for closeup]
Plot 4
scores: not implemented
data source:uves_fib_wave_scoreQC
(QC1 database)
dataset:resid_rms
median:1.98none
fixed thresholds:1.0...3.1none
N_data plotted:81
[click plot for closeup]
Plot 5
scores: 7|
data source:uves_fib_wave_fib5
(QC1 database)
dataset:resid_rms
median:2.03none
fixed thresholds:1.0...3.1none
N_data plotted:81
[click plot for closeup]
Plot 6
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resid_rms
median:1.99none
thresholds:none 
N_data plotted:81
[click plot for closeup]
Plot 7
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resid_rms
median:1.96none
thresholds:none 
N_data plotted:81
[click plot for closeup]
Plot 8
scores: not implemented
data source:uves_fib_wave
(QC1 database)
dataset:resid_rms
median:2.02none
thresholds:none 
N_data plotted:81
[click plot for closeup]
This plot

This is the trending report for FLAMES/UVES QC parameters derived for wavelength calibration. It covers the rms of the dispersion solution and is based on MOS data taken with the RED UPPER CCD in setting 580nm, 7+1FIB, with Plate 1. These data are taken daily as part of the Daily Health Checks.
Note: only Fibre 5 is considered in the score for this report.
The individual plots describe:

  • Fibre 1: the dispersion of the solution, measured as the rms of the residuals of the fitted wavelengths with respect to the laboratory wavelengths.
    Note that in contrast to fibres 2-8 which are illuminated by the robotic fibre projector fed by a Thorium-Argon lamp, fibre 1 is illuminated by the simultaneous calibration wavelength calibration lamp.
  • Fibres 2-8: the dispersion of the solution, measured as the rms of the residuals of the fitted wavelengths with respect to the laboratory wavelengths.

Note there is NO monitoring for fibre 9, used in the 8FIB modes.
Find more information about FLAMES/UVES wavelength solution monitoring here.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up