VIMOS trending system: HISTORY plot
Last update: 2011-07-10T03:29:22 (UT)
same group: QTH50_lamp He_lamp Ne_lamp Ar_lamp
close window history: 2009 2010   2011   2012   2013 2014 2015 2016 2017 2018   HEALTH
  ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEAN 3618826.5 ADU/s VAL2.5E6,4.5E6 165 flux this | last_yr | all Q2, flux from IFU flats
2 QC1DB MEAN 5195248.5 ADU/s VAL3.5E6,6.0E6 165 flux this | last_yr | all Q1, flux from IFU flats
3 QC1DB MEAN 4181356.75 ADU/s VAL2.5E6,5.0E6 165 flux this | last_yr | all Q3, flux from IFU flats
4 QC1DB MEAN 5635890.5 ADU/s VAL4.0E6,6.5E6 165 flux this | last_yr | all Q4, flux from IFU flats
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:vimos_ifu_score
(QC1 database)
dataset:flux
mean:3618826.5ADU/s
fixed thresholds:2500000.0...4500000.0ADU/s
N_data plotted:165
[click on plot for closeup]
Plot 2
data source:vimos_ifu_score
(QC1 database)
dataset:flux
mean:5195248.5ADU/s
fixed thresholds:3500000.0...6000000.0ADU/s
N_data plotted:165
[click on plot for closeup]
Plot 3
data source:vimos_ifu_score
(QC1 database)
dataset:flux
mean:4181356.75ADU/s
fixed thresholds:2500000.0...5000000.0ADU/s
N_data plotted:165
[click on plot for closeup]
Plot 4
data source:vimos_ifu_score
(QC1 database)
dataset:flux
mean:5635890.5ADU/s
fixed thresholds:4000000.0...6500000.0ADU/s
N_data plotted:165
[click on plot for closeup]
This plot

Trending plot for the QTH10 FF lamp.

The lamp performance is measured using IFU spectroscopic flats using the LR_red grism and the IFU shutter. The values should show smooth trending without sudden jumps. There are occasional pipeline failures in quadrant 2 that cause an artificially low value for the extracted flux. These are not due to instrument problems.

Template ID: ifu_cal_DayCalib: Pipeline recipe: vmifucalib.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up