VIMOS trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2018-04-04T03:38:29 (UT)
  | now: 2024-04-27T04:04:03 (UT) muc03 QC pipeline: vimos-2.9.19 (installed 2016-03-17)
same group: QTH50_lamp He_lamp Ne_lamp Ar_lamp
close window history: 2009 2010 2011 2012 2013 2014 2015 2016 2017   2018   HEALTH
  ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 3.41e+06 ADU/s VAL2.5E6,4.5E6 150 flux this | last_yr | all Q2, flux from IFU flats
2 QC1DB MEDIAN 5.25e+06 ADU/s VAL3.5E6,6.0E6 150 flux this | last_yr | all Q1, flux from IFU flats
3 QC1DB MEDIAN 4.16e+06 ADU/s VAL2.5E6,5.0E6 150 flux this | last_yr | all Q3, flux from IFU flats
4 QC1DB MEDIAN 5.45e+06 ADU/s VAL4.0E6,6.5E6 150 flux this | last_yr | all Q4, flux from IFU flats
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:vimos_ifu_score
(QC1 database)
dataset:flux
median:3.41e+06ADU/s
fixed thresholds:2500000.0...4500000.0ADU/s
N_data plotted:150
[click plot for closeup]
Plot 2
scores: not implemented
data source:vimos_ifu_score
(QC1 database)
dataset:flux
median:5.25e+06ADU/s
fixed thresholds:3500000.0...6000000.0ADU/s
N_data plotted:150
[click plot for closeup]
Plot 3
scores: not implemented
data source:vimos_ifu_score
(QC1 database)
dataset:flux
median:4.16e+06ADU/s
fixed thresholds:2500000.0...5000000.0ADU/s
N_data plotted:150
[click plot for closeup]
Plot 4
scores: not implemented
data source:vimos_ifu_score
(QC1 database)
dataset:flux
median:5.45e+06ADU/s
fixed thresholds:4000000.0...6500000.0ADU/s
N_data plotted:150
[click plot for closeup]
This plot

Trending plot for the QTH10 FF lamp.

The lamp performance is measured using IFU spectroscopic flats using the LR_red grism and the IFU shutter. The values should show smooth trending without sudden jumps. There are occasional pipeline failures in quadrant 2 that cause an artificially low value for the extracted flux. These are not due to instrument problems.

Template ID: ifu_cal_DayCalib: Pipeline recipe: vmifucalib.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up