VIRCAM trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2013-05-03T14:05:05 (UT)
same group: MED_1.0011 RON12_1.0011 RMS_1.0011 STRIPE_1.0011   MED_120 RON12_120 RMS_120 STRIPE_120
close window history: 2009 2010 2011 2012   2013   2014   2015 2016 2017 2018 2019 2020 2021 2022 2023   HEALTH FULL
  ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
  • (old) static REFERENCE frames
  • (new) dynamic REFERENCE frames
  • Scoring (red/green squares in the group bar and in the navigation bar) is on the detector average (Fig. 1) and detector rms (Fig. 3) values alone.
  • Scores of individual detector values (Fig. 4) are for detailed analysis but do not contribute to the overall instrument score.
  • All VIRCAM thresholds are based on the following two documents:
    • VIS-SPE-ATC-00000-0008, "VISTA Technical Verification Specification"
    • ESO-347147, "VIRCAM Crtitical Control Parameters"
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN -0.038 ADU VAL-1.5,2 72 qc_darkmed this | last_yr | all DARKMED, average over all 16 detectors
2 QC1DB MEDIAN -0.0811 ADU VAL-1.5,2 1145 qc_darkmed this | last_yr | all DARKMED, values of all detectors
3 QC1DB MEDIAN 0.351 ADU VAL-1.5,2 72 qc_darkmed this | last_yr | all DARKMED, stdev over all detectors
4 LOCAL none   ADU none  1152 qc_darkmed n/a Score per Detector: Green = within the thresholds
4 LOCAL none   ADU none  0 qc_darkmed n/a Score per Detector: Red = beyond thresholds
4 .LOCAL none   ADU none  1152 qc_darkmed n/a Score per Detector: Green = within the thresholds
4 .LOCAL none   ADU none  0 qc_darkmed n/a Score per Detector: Red = beyond thresholds
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:vircam_dark
(QC1 database)
dataset:qc_darkmed
median:-0.038ADU
fixed thresholds:-1.5...2ADU
N_data plotted:72
[click on plot for closeup]
Plot 2
data source:vircam_dark
(QC1 database)
dataset:qc_darkmed
median:-0.0811ADU
fixed thresholds:-1.5...2ADU
N_data plotted:1145
[click on plot for closeup]
Plot 3
data source:vircam_dark
(QC1 database)
dataset:qc_darkmed
median:0.351ADU
fixed thresholds:-1.5...2ADU
N_data plotted:72
[click on plot for closeup]
Plot 4
data source:local text file
dataset:
(numbers below apply to this dataset)
qc_darkmed
average:none
thresholds:none
N_data plotted:1152
[click on plot for closeup]
This plot

Trending plot for median dark level.

QC parameters is derived from each detector of daily DARK frames.

Name:
Fits header: QC.DARKMED,
QC1 DB: qc_darkmed

Description:
Median of the MASTER_DARK pipeline product frame

Setup:
DET.DIT = 1.0011 sec

DET.NDIT = 1

These plots describe:
Fig. 1. Average view: The QC parameter averaged over all detectors. This view is sensitive to common changes.
Fig. 2. Resolved View: The QC parameter resolved for all detectors.
Fig. 3. Scatter View: The standard deviation of the QC parameter, obtained from the sample of all detectors. This view is sensitive changes of individual detectors.
Fig. 4. Scoring View: Scoring values of all 16 QC parameters. Green = within the acceptance thresholds. Red = beyond the acceptance thresholds.

Find more information on the VIRCAM DARK tutorial page.

 


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up