Plot ? |
Symb ? |
Source * |
Average ? |
Thresholds ? |
N_ data |
QC1 parameter |
Data downloads |
Remarks |
method |
value |
unit |
method |
value |
1 |
| QC1DB |
MEDIAN |
2.89 |
ADU |
none | |
5388 |
qc_darkrms |
all (full time range)
| DARKRMS, average over all 16 detectors |
2 |
| QC1DB |
MEDIAN |
2.96 |
ADU |
none | |
84739 |
qc_darkrms |
all (full time range)
| DARKRMS, values of all detetcors |
2 |
| QC1DB |
MEDIAN |
3.07 |
ADU |
none | |
5362 |
qc_darkrms |
all (full time range)
| DARKRMS, values of detector 7 |
2 |
| QC1DB |
MEDIAN |
2.99 |
ADU |
none | |
5368 |
qc_darkrms |
all (full time range)
| DARKRMS, values of detector 3 |
3 |
| QC1DB |
MEDIAN |
0.335 |
ADU |
none | |
5389 |
qc_darkrms |
all (full time range)
| DARKRMS, stdev over all detectors |
|
*Data sources: QC1DB: QC1 database; LOCAL: local data source
|
Plot 1 scores: not implemented
| data source: | vircam_dark (QC1 database) |
dataset: | qc_darkrms | |
median: | 2.89 | ADU |
thresholds: | none | |
N_data plotted: | 5388 |
[click plot for closeup] |
Plot 2 scores: not implemented
| data source: | vircam_dark (QC1 database) |
dataset: (numbers below apply to this dataset) | qc_darkrms | |
median: | 2.96 | ADU |
thresholds: | none | |
N_data plotted: | 84739 |
[click plot for closeup] |
Plot 3 scores: not implemented
| data source: | vircam_dark (QC1 database) |
dataset: | qc_darkrms | |
median: | 0.335 | ADU |
thresholds: | none | |
N_data plotted: | 5389 |
[click plot for closeup] |
This plot
Trending plot for RMS noise of MASTER_DARK.
QC parameters is derived from each detector of daily DARK frames.
Name:
Fits header: QC.DARKRMS
QC1 DB: qc_darkrms
Description:
RMS noise of the MASTER_DARK pipeline product frame
Setup:
DET.DIT = 1.0011 sec
DET.NDIT= 1
These plots describe:
Fig. 1. Average view: The QC parameter averaged over all detectors.
This view is sensitive to common changes.
Fig. 2. Resolved View: The QC parameter resolved for all detectors.
Fig. 3. Scatter View: The standard deviation of the QC parameter, obtained from the sample of all detectors.
This view is sensitive changes of individual detectors.
Fig. 4. Scoring View: Scoring values of all 16 QC parameters.
Green = within the acceptance thresholds.
Red = beyond the acceptance thresholds.
Find more information on the VIRCAM DARK tutorial page.
General information
Click on any of the plots to see a close-up version.
The latest date is indicated on top of the plot.
If configured,
- data points belonging to the latest date are specially marked
- statistical averages are indicated by a solid line, and thresholds by broken lines
- outliers are marked by a red asterisk. They are defined as data points outside the
threshold lines
- "aliens" (= data points outside the plot Y limits) are marked by a red arrow (↑ or ↓)
- you can download the data for each parameter set if the 'Data downloads' link shows up
|