VISIR trending system: HISTORY plot
Last update: 2010-10-15T02:02:35 (UT)
same group: read-noise dark det-temp
close window history: 2006 2007 2008 2009   2010   2011   2012 2013 2014 2015   HEALTH
  ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEAN 118.225   VAL100,150 80 nbbadpix this | last_yr | all Number of bad pixels - imaging detector
2 LOCAL MEAN 48.4125   VAL40,70 80 badp_splog n/a Number of bad pixels - spectroscopic detector
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:visir_flat_img
(QC1 database)
parameter | dataset:nbbadpix
mean:118.225 
fixed thresholds:100.0...150.0 
N_data:80
[click on plot for closeup]
Plot 2
data source:local text file
parameter | dataset:badp_splog
mean:48.4125 
fixed thresholds:40.0...70.0 
N_data:80
[click on plot for closeup]
This plot

This is a trending plot monitoring the number of bad pixels as calculated by the pipeline from the internal flat field frames. The bad pixel values are displayed for both, the imaging and spectroscopic detectors.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up