VISIR trending system: HISTORY report
Last update: 2012-04-15T02:03:53 (UT)
same group: read-noise dark det-temp
close window history: 2006 2007 2008 2009 2010 2011   2012   2013   2014 2015   HEALTH
  ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEAN 57.54298 NONE VAL45,70 94 ron2 this | last_yr | all read-out-noise of the imaging detector
2 QC1DB MEAN 20.13836 NONE VAL0,70 186 ron2 this | last_yr | all read-out-noise of the spectroscopic detector
*Data sources: QC1DB: QC1 database; LOCAL: local text file
Plot 1
data source:visir_dark_img
(QC1 database)
dataset:ron2
mean:57.54298NONE
fixed thresholds:45.0...70.0NONE
N_data plotted:94
[click on plot for closeup]
Plot 2
data source:visir_dark_spc
(QC1 database)
dataset:ron2
mean:20.13836NONE
fixed thresholds:0.0...70.0NONE
N_data plotted:186
[click on plot for closeup]
This plot

Trending plot monitoring the read-out-noise (RON) measured from the raw(3)-raw(2) difference frame.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up