[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+R to enforce refresh of scores and dates]
VISIR trending system: HEALTH CHECK report
Last update: 2012-05-26T02:04:57 (UT)
| now: 2012-05-26T09:08:26 (UT)
HELP USERS-GUIDE MORE
same group: read-noise read-noise12 dark flat-flux nbadpix det-temp
General news:
VISIR news: 2012-05-09: VISIR upgrade began; instrument is scheduled back in operations on Sep.9th, 2012
Report news:  
DATE*: [?] 2012-05-19
no data | NR
2012-05-20
no data | NR
2012-05-21
no data | NR
2012-05-22
no data | NR
2012-05-23
no data | NR
2012-05-24
no data | NR
2012-05-25
no data | NR
P...
Product quality:
no products no products no products no products no products no products no products
  last OPSLOG data:
2012-05-08*

last files ...
monitor
close window history: 2006 2007 2008 2009 2010 2011   2012   HEALTH
  ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (0) MEAN 5.98304 K VAL5.8,6.1 70 ins_sens30_val this | last_yr | all temperature of the imaging detector
2 QC1DB yes (0) MEAN 7.25558 K VAL6.8,7.5 137 ins_sens31_val this | last_yr | all temperature of the spectroscopic detector
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
Plot 1
scores: 2|
data source:visir_dark_img
(QC1 database)
dataset:ins_sens30_val
mean:5.98304K
fixed thresholds:5.8...6.1K
N_data plotted:70
[click on plot for closeup]
Plot 2
scores: 4|
data source:visir_dark_spc
(QC1 database)
dataset:ins_sens31_val
mean:7.25558K
fixed thresholds:6.8...7.5K
N_data plotted:137
[click on plot for closeup]
This plot

This is a trending plot monitoring temperatures of the VISIR imaging and spectroscopic detectors. As of May 12, 2010 the temperature of the imaging detector is read from keyword INS_TEMP30_VAL and of the spectroscopic detector from keyword INS_TEMP31_VAL. Before that date different keywords were used - INS_SENS30_VAL and INS_SENS31_VAL respectively.  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPS column.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up