setup used for monitoring is: 100k/1pt/hg read mode, 1.0x11 slit and 1x1 binning
D2 lamp is monitored via flux1 QC parameters
QTH lamp is monitored via flux2 QC parameters
Arguments for upper threshold of 60000 ADU:
- The flat field lamp fluctuations within a set of five raw frames can be +- 1000 ADU.
- When the counts QC parameter reaches 63000 ADU the first hot pixel begin to saturate.
- The UVB-arm detector is linear ( < 0.01) up to the saturation level of 65535 ADU.