According to
- ESO-327242, Integrating Instrument Performance in the Quality Control Process - X-shooter
- VLT-SPE-ESO-14650-3323, X-SHOOTER Technical Specification
The read out noise (RON) is specified to be
- UVB/VIS low gain: <3e-
- UVB/VIS high gain: <8e-
- NIR (single read out): <15e-
- NIR (multiple nondestructive read out): <5e-
Given the monitored gain values of
- UVB low gain: 1.84 [e-/ADU]
- UVB high gain: 0.64 [e-/ADU]
- VIS low gain: 1.66 [e-/ADU]
- VIS high gain: 0.64 [e-/ADU]
- NIR 2.29 [e-/ADU]
the following KPI plot reference values can be derived:
- UVB low gain: 4.3 ADU
- UVB high gain: 4.7 ADU
- VIS low gain: 4.8 ADU
- VIS high gain: 4.7 ADU
- NIR (single readout): 6.5 ADU
- NIR (multiple nondestructive readout): 2.18 ADU