| GIRAFFE Quality Control:
wavelength solution |
|||||
|
|
The quality of the dispersion solution depends on the quality of the arclamp exposure and of the pipeline recipe. The arclamp exposure is quality-controlled in the arclamp plot. The results of the pipeline recipe are monitored here, in terms of three parameters:
The arc lamp data are measured in a pattern involving two health check settings: the high-resolution H525.8B setup, and the low-resolution L543.1 setup. The former is measured in every slit system, the latter only in the two Medusa's. The pattern repeats every week. QC1 parameters The QC1 parameter rms_fit (PRO.WSOL.RMS) describes the rms of the residuals of the accepted line positions against the dispersion solution.
Trending The trending plot monitors the HC data taken in high resolution (H525.8B, upper panel) and low resolution (L543.1, Medusa1 and 2 only, lower panel). History
This parameter is derived by the pipeline from the same set of Health Check data as the residual RMS: arclamp file in high resolution (H525.8B) and low resolution (L543.1, Medusa1 and 2 only). QC1 parameter
Note that the ops log data contain only the parameter PRO.WSOL.NLINES which measures the nominal number of lines per fibre, before acceptance. This value is usually slightly higher than nlin_acc. Trending
Plots 6-7 show the results for the HC settings for the low resolution grating:
For the other settings, this parameter varies strongly. Therefore it is not possible to provde a trending for each individual setup, but there are three plots which monitor the chromatic dependence of nlin_acc:
History
The resolution of the GIRAFFE arclamp calibration data is currently monitored by measuring the width of the identified and accepted arc emission lines. The QC1 parameter mean_resol in pixels describes the fibre-averaged width of the emission lines. There are currently no ops log data for this parameter. QC1 parameters
Trending The upper panel of the trending plot monitors the HC data taken in high resolution (H525.8B, all five fibre systems), the lower panel those for low resolution (L543.1, Medusa1 and 2 only). History
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]() |
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||