|
|
 |
| |
| QC PLOTS |
| |
CURRENT |
HISTORY |
| read-noise |
 |
 |
| ron |
 |
 |
| dark |
 |
 |
| dark_DCR |
 |
 |
| dark_FNS |
 |
 |
| dark_UCR |
 |
 |
| hot/cold pixels |
 |
 |
| QC1 database (advanced users):
browse
| plot |
Click on CURRENT to see the current trending (Health Check).
Click on HISTORY to see the historical evolution of the trending. |
A number of DARK frames is taken every day
within the NACO daytime calibrations. Each DARK consists of 3
raw frames obtained with the instrumental setting matching the
one used for science observations. The parameters matched are:
- DET.DIT - detector integration time
- DET.NDIT - number of sub-integrations
- INS.OPTI7.NAME - camera
- DET.NCORRS.NAME - detector read-out mode
- DET.MODE.NAME - detector mode
In addition, each day, 3 dark frames are taken for the health check purpose, always with the same setup:
- DET.DIT=1.7927 s, DET.NDIT=1, INS.OPTI7.NAME=S27, DET.NCORRS.NAME=FowlerNsamp, DET.MODE.NAME=HighSensitivity
- DET.DIT=0.5 s, DET.NDIT=100, INS.OPTI7.NAME=S27, DET.NCORRS.NAME=Double_RdRstRd, DET.MODE.NAME=HighSensitivity
- DET.DIT=300 s, DET.NDIT=1, INS.OPTI7.NAME=S27, DET.NCORRS.NAME=FowlerNsamp, DET.MODE.NAME=HighSensitivity
They are utilized to monitor performance of the NACO detector.
 |
DARK frame - 0.5s, Double_RdRstRd, HighSensitivity
This is an example of the master calibration DARK frame -
NC_DCAL_0_5_S27_F_DRHS.fits. It is a pipeline product of the
"naco_img_dark" recipe. The horizontal stripes on the edges of the
detector are features characteristic to the Double_RdRstRd read-out
mode. |
 |
DARK frame - 300s, FowlerNsamp, HighSensitivity
This is an example of the master calibration DARK frame -
NC_DCAL_300_S27_F_FSHS.fits. It is a pipeline product of the
"naco_img_dark" recipe. The bright, diagonal stripes are associated
with the "light leakage". They become prominent structures when darks
are taken with long DITs. |
QC1 parameters
| parameter |
QC1 database: table, name |
procedure |
| "dark_ron12": read-out noise |
naco_dark, dark_ron12 |
read-out noise calculated from the 2nd-1st raw; it is a median of the standard deviations measured from 100 4x4 windows on the difference frame, multiplied by sqrt(NDIT/2) |
The following plot shows the temporal evolution of the
read-out noise for different detector read-out modes:
DoubleReadResetRead (panel 1), Uncorrelated (panel 2),
FowlerNSample, DIT <= 90s (panel 3), and FowlerNSample, DIT >
90s (panel 4). In case of the FowlerNSample mode the read-out
noise was found to depend strongly on the DIT. Thus, it is
separated for the short and long DITs.
The other plot combines all the read-out noise
values separated by the detector read-out mode.
QC1 parameters
| parameter |
QC1 database: table, name |
procedure |
| "dark_median": median dark level (ADU/pixel/s) |
naco_dark, dark_median |
median dark level in the central part of the master dark |
| "dark_min": minimum of the dark level (ADU/pixel/s) |
naco_dark, dark_min |
calculated by the stcube function |
| "dark_max": maximum of the dark level (ADU/pixel/s) |
naco_dark, dark_max |
calculated by the stcube function |
The health check
plot shows the dark current median value measured for the daily darks taken with DIT=300s,
DET.NCORRS.NAME=FowlerNSamp, DET.MODE.NAME=HighSensitivity (plot 1), and with DET.DIT=0.5s, DET.NCORRS.NAME=Double_RdRstRd,
DET.MODE.NAME=HighSensitivity (plot 2).
Trending
The additional trending plots show the dark current median calculated for the darks taken with:
- the DoubleRdRstRd read-out mode dark_DCR,
- the FowlerNSamp read-out mode dark_FNS,
and
- the Uncorrelated read-out mode dark_UCR
QC1 parameters
| parameter |
QC1 database: table, name |
procedure |
| "num_cold": number of cold pixels |
naco_dark, num_cold |
calculated by the pipeline recipe naco_img_dark |
| "num_hot": number of hot pixels |
naco_dark, num_hot |
calculated by the pipeline recipe naco_img_dark |
The trending
plot shows the number of the hot and cold pixels measured in the dark frames. The data points
are separated by the detector read-out mode.
|