We
show the most critical QC parameters, derived from the SINFONI pipeline sinfo_rec_wavecal
recipe, measuring performance characteristics of the spectroscopic components
and wavelength calibration.
Two data set are shown: QC parameter derived
by the pipeline in Garching (QC) and QC parameters derived by the pipeline on
Paranal (PSO). In addition, the average of the Garching data shown. In case no
Garching data are available the average of the PSO data is given instead.
The
QC parameter definitions can be found on the qc
wave page. Since there is a wavelength solution of nearly all of the more
than 2000 columns of the detector, the average of the coefficients is trended.
When no QC parameters are available for a setup, the average is set to
0.
The three boxes per plot corresponds the three cameras=0.25, 0.1 and
0.025.
Grating offset (the pipeline central
wavelength result minus the nominal value (e.g. 2.200 micron for K, 1.6500 micron
for H ...)). A plot contains for each of the three cameras ( =
0.025, 0.1, 0.25) a box
The blue points are
processed on Paranal ( = PSO)
The black points are processed
by QC Garching ( = QC)