Offered spectroscopic modes
Spectroscopy can be performed using one of 4 offered grisms or a prism.
The resolution quoted here have been computed for the 86mas slit.
The spectral resolution for slitless spectroscopy and spectrum taken with the 172mas slit is set by the PSF.
Prism spectroscopy is not offered on NACO since P84. The information on the prism is kept for archival data only.
Arc spectra with the long wavelength modes are not possible.
| Mode Name | Camera Name | Order | Spectral Range | Linear Dispersion | Resolution | Arc Atlas |
| S54_4_SJ | S54 | 1 | 0.91-1.40 | 2.00 | 400 | S54_4_SJ_wc.pdf |
| S54_3_SH | S54 | 3 | 1.37-1.84 | 0.69 | 1500 | S54_3_SH_wc.pdf |
| S27_3_SH | S27 | 3 | 1.37-1.72 | 0.34 | 1500 | Not available |
| S27_4_SH | S27 | 1 | 1.37-1.84 | 0.97 | 500 | S27_4_SH_wc.pdf |
| S54_2_SK | S54 | 2 | 1.79-2.42 | 0.97 | 1400 | S54_2_SK_wc.pdf |
| S27_2_SK | S27 | 2 | 1.79-2.24 | 0.59 | 1400 | S27_2_SK_wc.pdf |
| S54_3_SK | S54 | 2 | 1.79-2.57 | 1.00 | 1400 | S54_3_SK_wc.pdf |
| S27_3_SK | S27 | 2 | 2.02-2.53 | 0.50 | 1400 | S27_3_SK_wc.pdf |
| S54_4_SK | S54 | 1 | 1.79-2.57 | 1.96 | 700 | S54_4_SK_wc.pdf |
| L27_1_L | L27 | 2 | 3.20-3.76 | 1.60 | 700 | Not available |
| L54_2_L | L54 | 1 | 3.20-3.76 | 2.00 | 1100 | Not available |
| L27_1_LP | L27 | 2 | 3.50-4.10 | 1.60 | 700 | Not available |
| L54_2_LP | L54 | 1 | 3.50-4.10 | 2.00 | 1100 | Not available |
| L27_2_LP | L27 | 1 | 3.50-4.10 | 1.00 | 1100 | Not available |
| S54_4_SHK | S54 | 1 | 1.30-2.60 | 1.94 | 550 | S54_4_SHK.pdf |
| L54_1_SL | L54 | 2 | 2.60-4.20 | 3.16 | 700 | Not available |
| L27_1_SL | L27 | 2 | 2.60-4.10 | 1.57 | 700 | Not available |
| L54_2_SL | L54 | 1 | 3.02-4.20 | 2.01 | 1100 | Not available |
| L27_2_SL | L27 | 1 | 3.47-4.20 | 1.00 | 1100 | Not available |
List of Argon Lines in calibration arcs
| Mode Name | Filter | Dispersion (nm/pixel) |
Fit validity (μm) |
Wavelength range (μm) | Resolution | Fit order | Fit rms (nm) |
| L27_P1 | None | 8.5 | 1 - 4 | 0.85 - 5.5 | 90 | 3 | 10 |
| L27_P1 | None | 6.3 | 4.5 - 5.5 | 0.85 - 5.5 | 250 | 5 | 2.9 |
| S13_P1 | CutOff_2.5um | 4.1 | - | 0.85 - 2.5 | 60 | 3 | 10 |
| S27_P1 | CutOff_2.5um | 8.2 | - | 0.85 - 2.5 | 60 | - | - |
Transmission curve for the Cutoff 2.5 micron filter used to select the not-thermal wavelength range with the prims.
In order to get an idea of the performance of this mode, an example spectrum (DIT=0.5, 86mas slit) of a red 2MASS star with J=8.04, H=7.20 and K=7.01 is provided (courtesy C. Lidman).
Raw data in two nod positions:
Spectrum nod A
Spectrum nod B
Subtracted frame: Spectrum: nod A - nod B