| This documentation is intended both for QC scientists and SciOps astronomers (who may want to ignore the technical information displayed in grey).
|
| |
|
|
|
|
| |
|
bottom
plots
|
contact
|
|
| NAME |
qc_FF_SPEC.sh
|
|
| VERSION |
1.0 -- created (2005-03-07)
1.1 -- QC parameters extracted from headers of products rather than paf files (2005-11-07)
|
|
| SYNTAX |
SHELL
|
|
| CALL |
processQC
from $DFS_PRODUCT/FF_SPEC/$DATE:
qc_FF_SPEC $date $primfile "
|
|
| INSTRUMENT |
VISIR
|
|
| RAWTYPE |
FF_SPEC
|
|
| PURPOSE |
extracts QC parameters and records them in the local qclog table $DFO_TREND_DIR/FF_SPEC.qclog as well as ingests in the qc1 database, table: visir_flat_spc
|
|
| PROCINPUT |
spectral flatfield product
|
|
| QC1TABLE |
trending | table(s) in QC1 database:
visir_flat_spc
|
|
| TRENDPLOT |
trending | HealthCheck plot(s) associated to this procedure:
trend_report_BADPIX_HC.html
|
|
| QC1PAGE |
trending | associated documentation: none
|
|
QC1PLOTS
top |
|
|
| QC1PARAM |
QC1 parameters written into local QC1 table $DFO_TREND_DIR/FF_SPEC.qclog:
QC.NBBADPIX
FITS key names: INS.PFOV | INS.FILT2.NAME | INS.SLIT1.WID | INS.RESOL | INS.GRAT1.WLEN
QC1 params created by pipeline: all
QC1 params created by this procedure: none
|
|
| ALGORITHM |
Description of algorithms:
For detailed description of algorithms used by the recipe see the VISIR Pipeline User Manual
QC.NBBADPIX: for each pixel gain is calculated (slope of the pixel intensity vs. median); pixels with relative gain: 0.2 < gain < 1.5 are flagged as bad
|
|
| CERTIF |
Reasons for rejection: - NONE; product is not used for data calibration
|
|
| COMMENTS |
NONE
|
| top |
[2012-06-20T10:56:59] created by qcDocu v1.1.2, a tqs tool
|