| This documentation is intended both for QC scientists and SciOps astronomers (who may want to ignore the technical information displayed in grey).
|
| |
|
|
|
|
| |
|
bottom
plots
|
contact
|
|
| NAME |
qc_STD_IMG.sh
|
|
| VERSION |
1.0 -- created (2005-03-07)
1.1 -- QC parameters extracted from headers of products (2005-11-07)
1.2 -- extended the list of QC parameters (2006-09-26)
|
|
| SYNTAX |
SHELL
|
|
| CALL |
measureQuality
from $DFS_PRODUCT/STD_IMG/$DATE:
qc_STD_IMG $date $primfile "
|
|
| INSTRUMENT |
VISIR
|
|
| RAWTYPE |
STD_IMG
|
|
| PURPOSE |
extracts QC parameters and records them in the local qclog table $DFO_TREND_DIR/STD_IMG.qclog and in qc1 database table visir_zp_img
|
|
| PROCINPUT |
imaging standard star product
|
|
| QC1TABLE |
trending | table(s) in QC1 database:
visir_zp_img
|
|
| TRENDPLOT |
trending | HealthCheck plot(s) associated to this procedure:
trend_report_SENS_HC.html
trend_report_NSENS_HC.html
trend_report_QSENS_HC.html
trend_report_CONVER_HC.html
trend_report_BACKGD_HC.html
|
|
| QC1PAGE |
trending | associated documentation: none
|
|
QC1PLOTS
top |
|
|
| QC1PARAM |
QC1 parameters written into QC1 table: sensit | convert | strehl | strehl_error | backgd_mean/sigma | fwhm
QC1 parameters written into local QC1 table $DFO_TREND_DIR/STD_IMG.qclog:
QC.SENSIT | QC.CONVER | QC.STREHL | QC.STREHL.ERROR | QC.BACKGD.MEAN | QC.JYVAL | QC.FWHMX | QC.FWHMY
QC.SENSIT: - computed in units mJy at 10sigma in 1 houry
FITS key names: INS.PFOV | DET.DIT | FILTER1 | AIRMASS
QC1 params created by pipeline: all
QC1 params created by this procedure: none
|
|
| ALGORITHM |
Description of algorithms:
For details see VISIR Pipeline User Manual<
|
|
| CERTIF |
Reasons for rejection: - target too faint
Reasons for rejection: - detector artifacts too bright
Frame is displayed to check for unusual patterns, artificial structures
|
|
| COMMENTS |
NONE
|
| top |
[2012-06-20T10:57:04] created by qcDocu v1.1.2, a tqs tool
|