Plot 1
data source:
fors2_mos_wave (QC1 database)
dataset:
central_resol
average:
none
thresholds:
none
N_data plotted:
1163
(current) reference:
503
[click plot for closeup]
Plot 2
data source:
fors2_mos_wave (QC1 database)
dataset:
central_resol
average:
none
thresholds:
none
N_data plotted:
1160
(current) reference:
508
[click plot for closeup]
This plot
This is the instrument performance plot for the FORS2 MOS resolution.
From a wavelength calibrated MOS arc lamp exposure, the mean spectral
resolution from all lines in catalog is computed. Each single spectral
resolution is computed as the ratio between the arc lamp line wavelength
and its FWHM. The values trended here are for the 300V grism with a slit
width of 1.0" and the order separation filter GG435.
Data have been acquired using the red-sensitive MIT detector.
General information
Click on any of the plots to see a close-up version.
The latest date is indicated on top of the plot.
If configured,
data points belonging to the latest date are specially marked
statistical averages are indicated by a solid line, and thresholds by broken lines
outliers are marked by a red asterisk. They are defined as data points outside the
threshold lines
"aliens" (= data points outside the plot Y limits) are marked by a red arrow (↑ or ↓)
you can download the data for each parameter set if the 'Data downloads' link shows up