mirror sites: PL (internal link) HQ     [?]

Quality Control and
Data Processing


[an error occurred while processing this directive] SPHERE: ZIMPOL dark frames
 
HC PLOTS
dark level IMG readout mode [ADU/h]
dark level POL slow read mode CAM1 [ADU/h]
dark level POL slow read mode CAM2 [ADU/h]
dark level POL fast read mode CAM1 [ADU/h]
dark level POL fast read mode CAM1 [ADU/h]
QC1 database (advanced users): browse | plot
QC1 database (advanced users): browse | plot

Optical dark frames are acquired once a month in all three read modes to monitor the dark radiation. The exposure time is 100 sec, the QC paramaters in the plot are scaled to ADU/h.


ZIMPL Dark Level

QC1_parameters

FITS key QC1 database: table, name definition class* HC_plot** more docu
QC.DOUBLE.IMAGE.IFRAME.MEDIAN sphere_zimpol_dark_img..qc_median_iframe HC [docuSys coming]
QC.DOUBLE.IMAGE.LEAKAGE.MEDIAN sphere_zimpol_dark_img..qc_median_pframe median counts polarimetric HC [docuSys coming]
QC.DOUBLE.IMAGE.IFRAME.RMS sphere_zimpol_dark_img..qc_rms_iframe rms of counts imageHC [docuSys coming]
QC.DOUBLE.IMAGE.LEAKAGE.RMS sphere_zimpol_dark_img..qc_rms_pframe rms of counts polarimetric HC [docuSys coming]
QC.QUAD.IMAGE.ZERO.ODD.MEDIAN sphere_zimpol_dark_pol..qc_med_zero_odd polphase zero, odd cycle HC [docuSys coming]
QC.QUAD.IMAGE.ZERO.EVEN.MEDIAN sphere_zimpol_dark_pol..qc_med_zero_even polphase zero, even cycleHC [docuSys coming]
QC.QUAD.IMAGE.PI.ODD.MEDIAN sphere_zimpol_dark_pol..qc_med_pi_odd polphase pi, odd cycleHC [docuSys coming]
QC.QUAD.IMAGE.PI.EVEN.MEDIAN sphere_zimpol_dark_pol..qc_med_pi_even polphase pi, even cycle HC [docuSys coming]
QC.QUAD.IMAGE.ZERO.ODD.RMS sphere_zimpol_dark_pol..qc_rms_zero_odd polphase zero, odd cycle HC [docuSys coming]
QC.QUAD.IMAGE.ZERO.EVEN.RMS sphere_zimpol_dark_pol..qc_rms_zero_even polphase zero, even cycleHC [docuSys coming]
QC.QUAD.IMAGE.PI.ODD.RMS sphere_zimpol_dark_pol..qc_rms_pi_odd polphase pi, odd cycleHC [docuSys coming]
QC.QUAD.IMAGE.PI.EVEN.RMS sphere_zimpol_dark_pol..qc_rms_pi_even polphase pi, even cycle HC [docuSys coming]
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one.

Trending

The plot shows the optical CCD dark current values and RMS as derived from the zimpol dark recipe for read mode StandardImaging. The recipe expresses dark values in units of the used EXPTIME (here per 100 sec), the values in the plot are scaled to ADU/hour.

The plot shows QC parameters derived from the ZIMPOL dark recipe for both read modes SlowPolarimetry and FastPolarimetry.

odd/even means the cycle
zero/pi means the polarimetric phase

The recipe arranges the phase=pi rows and the phase=zero rows of the raw frame and writes them in different extensions of the pipeline product. The QC parameters describe the median and the rms of these imaging products. The QC parameter 'median' retrieved and stored is given in ADU per 100 sec, the values in the plots are scaled to ADU/hour.

Scoring&thresholds ZIMPL Dark Level

The median dark levels are scored.

History

  • Due to a contamination issue dark values in 2015 are higher.
  • As occured on 2016-09-23, the saturated modemeff calibration acquired directly in advance of the ZIMPOL imaging mode dark calibration left a double as large dark current on the detector, a kind of persistence.

Algorithm ZIMPL Dark Level

The monitored values are directly derived from the intermediate pipeline product frames.


Send comments to <qc_sphere@eso.org> powered by QC [webCMS v1.0]