According to

The IRDIS detector read out noise is specified to be lower than 15e-. Given the gain (see user manual) of 1.75 e-/ADU, this results in 8.75 ADU.

Please note the temperature dependence documented in the FULL plots.

Evaluation:

For IRDIS DIT=2 and DIT=30, ron is below the specification. As the KPI plot reference value we choose the long-term average for data processed with the sphere pipeline recipe (after 2017): 4.62 ADU (DIT=2) and 3.06 ADU (DIT=30)