VIRCAM trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2023-06-12T10:26:05 (UT)
  | now: 2024-04-26T21:25:37 (UT) muc07 QC pipeline: vcam-2.3.9 (installed 2022-03-17)
same group: RESETMED RESETRMS RESETDIFFRMS   temperatures
close window history: 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022   2023   HEALTH
  ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 45.3 ADU VAL1,50 31 qc_resetdiff_rms this | last_yr | all RESETDIFFRMS, average over all 16 detectors
2 QC1DB MEDIAN 35.8 ADU VAL1,130 496 qc_resetdiff_rms this | last_yr | all RESETDIFFRMS, values of all detetcors
2 QC1DB none   ADU none  31 qc_resetdiff_rms this | last_yr | all RESETDIFFRMS, values of detector 2
2 QC1DB none   ADU none  31 qc_resetdiff_rms this | last_yr | all RESETDIFFRMS, values of detector 3
2 QC1DB none   ADU none  31 qc_resetdiff_rms this | last_yr | all RESETDIFFRMS, values of detector 13
3 QC1DB MEDIAN 39.6 ADU VAL0.1,50 31 qc_resetdiff_rms this | last_yr | all RESETDIFFRMS, stdev over all detectors
4 LOCAL none   ADU none  465 qc_resetdiff_rms n/a Score per Detector: Green = within the thresholds
4 LOCAL none   ADU none  31 qc_resetdiff_rms n/a Score per Detector: Red = beyond thresholds
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 3|
data source:vircam_reset
(QC1 database)
dataset:qc_resetdiff_rms
median:45.3ADU
fixed thresholds:1...50ADU
N_data plotted:31
[click plot for closeup]
Plot 2
scores: not implemented
data source:vircam_reset
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_resetdiff_rms
median:35.8ADU
fixed thresholds:1...130ADU
N_data plotted:496
[click plot for closeup]
Plot 3
scores: 3|
data source:vircam_reset
(QC1 database)
dataset:qc_resetdiff_rms
median:39.6ADU
fixed thresholds:0.1...50ADU
N_data plotted:31
[click plot for closeup]
Plot 4
scores: not implemented
data source:local data source
dataset:
(numbers below apply to this dataset)
qc_resetdiff_rms
average:none
thresholds:none
N_data plotted:465
[click plot for closeup]
This plot

Trending plot for RMS noise in MASTER_BIAS minus REFERENCE_BIAS combined product frame.

QC parameters is derived from each detector of daily RESET calibrations.

Name:
Fits header: QC.RESETDIFFRMS,
QC1 DB: qc_resetdiff_rms

Description:
RMS noise of master minus reference product frame

Setup:
DET.DIT = MINDIT = 1.0011 sec

These plots describe:
Fig. 1. Average view: The QC parameter averaged over all detectors. This view is sensitive to common changes.
Fig. 2. Resolved View: The QC parameter resolved for all detectors.
Fig. 3. Scatter View: The standard deviation of the QC parameter, obtained from the sample of all detectors. This view is sensitive changes of individual detectors.
Fig. 4. Scoring View: Scoring values of all 16 QC parameters. Green = within the acceptance thresholds. Red = beyond the acceptance thresholds.

Find more information on the VIRCAM RESET tutorial page.

 


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up