ESPRESSO: Bias
| |
| HC PLOTS |
| Bias level |
 |
| Bias RON |
 |
|
QC1 database (advanced users):
browse |
plot
|
Bias frames are measured during daytime. They are measured daily with 1x1 binning (fast read-out, for HR and UHR observations) and with
2x1 binning (slow read-out, for HR observations). Bias frames with 4x2 and 8x4 binning (slow read-out) are typically
only taken when MR science data have been observed during the night.
A complete set per binning consists of 10 bias frames.
 |
Bias frame. Example frame for the blue detector. Note the different bias level for the 16 read-out ports
|
Bias level
The master bias is the median stack of the (typically) ten input raw bias frames.
QC1_parameters
| FITS key |
QC1 database: table, name |
definition |
class* |
HC_plot** |
more docu |
| none | espresso_bias..bias_mean_avg | average bias level across read-out ports | CAL | | [docuSys coming] |
| none | espresso_bias..bias_mean_rms | RMS of bias level across read-out ports | CAL | | [docuSys coming] |
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one. |
ESPRESSO has two chips, one for the BLUE and one for the RED arm. Each chip
has 16 read-out ports. Bias QC parameters are determined by the pipeline per read-out port which
leads to 16 different values for each QC parameter per chip. They are written
as header keys in the form of (for the mean bias):
QC.EXT<n>.ROX<x>.ROY<y>.BIAS.MEAN
with <n> being the detector (0 for BLUE and 1 for RED),
<x> the x position of the
read-out on the chip (0, 1, to 7), and <Y> the y position (0 or 1). The
corresponding names in the QC1 database follow the schema
bias_mean_<x>_<y> .
In addition to these individual values, the average and the RMS of these
values are calculated per chip and written into the QC1 database.
Trending
The bias QC1 parameters are trended separately for the four different read-out
modes: 1x1 binning (fast read-out), and 2x1, 4x2, and 8x4 binning (all slow
read-out).
Scoring&thresholds Bias level
Thresholds have been set provisionally to +/- 25 ADUs during commissioning.
History
Bias level changed a few times during commisioning.
Algorithm Bias level
The mean bias level is determined per read-out port. The QC procedures calculate in addition the average and the RMS
of all 16 ports per detector.
Read-out noise (RON)
QC1_parameters
| FITS key |
QC1 database: table, name |
definition |
class* |
HC_plot** |
more docu |
| none | espresso_bias..bias_ron_avg | average RON across read-out ports | HC | | [docuSys coming] |
| none | espresso_bias..bias_ron_rms | RMS of RON across read-out ports | HC | | [docuSys coming] |
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one. |
ESPRESSO has two chips, one for the BLUE and one for the RED arm. Each chip
has 16 read-out ports. Bias QC parameters are determined by the pipeline per read-out port which
leads to 16 different values for each QC parameter per chip. They are written
as header keys in the form of (for the mean bias):
QC.EXT<n>.ROX<x>.ROY<y>.BIAS.MEAN
with <n> being the detector (0 for BLUE and 1 for RED),
<x> the x position of the
read-out on the chip (0, 1, to 7), and <Y> the y position (0 or 1). The
corresponding names in the QC1 database follow the schema
bias_mean_<x>_<y> .
In addition to these individual values, the average and the RMS of these
values are calculated per chip and written into the QC1 database.
Trending
The bias QC1 parameters are trended separately for the four different read-out
modes: 1x1 binning (fast read-out), and 2x1, 4x2, and 8x4 binning (all slow
read-out).
Scoring&thresholds Read-out noise (RON)
TBD.
History
There was a period with increased electronic noise during commisioning.
Algorithm Read-out noise (RON)
RON is determined per read-out port. The QC procedures calculate in addition the average and the RMS
of all 16 ports per detector.
|