[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
FORS2 trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-04-29T17:21:19 (UT)
  | now: 2024-04-29T18:39:14 (UT) muc01 QC pipeline: fors-5.6.10 (installed 2024-04-19)
same group: med_master ron_raw ron_master fp_raw struct_raw pattern
close window history: 2006 2007 2008 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023   2024   HEALTH
  ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
2014-11-26: 20 rather than 5 input frames are now taken in the BIAS template; This lowers the read-out-noise level by a factor of sqrt(20/5)=2; Thresholds adjusted
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 0.93 ADU VAL0.6,1.3 80 noise_master this | last_yr | all Chip 1, 100Kps/2ports/high_gain, 2x2
2 QC1DB MEDIAN 1.12 ADU VAL0.6,1.3 80 noise_master this | last_yr | all Chip 2, 100Kps/2ports/high_gain, 2x2
3 QC1DB MEDIAN 0.657 ADU VAL0.2,1.0 87 noise_master this | last_yr | all Chip 1, 200Kps/2ports/low_gain, 2x2
4 QC1DB MEDIAN 0.725 ADU VAL0.2,1.0 87 noise_master this | last_yr | all Chip 2, 200Kps/2ports/low_gain, 2x2
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 2|
data source:fors2_bias
(QC1 database)
dataset:noise_master
median:0.93ADU
fixed thresholds:0.6...1.3ADU
N_data plotted:80
[click plot for closeup]
Plot 2
scores: 2|
data source:fors2_bias
(QC1 database)
dataset:noise_master
median:1.12ADU
fixed thresholds:0.6...1.3ADU
N_data plotted:80
[click plot for closeup]
Plot 3
scores: 4|
data source:fors2_bias
(QC1 database)
dataset:noise_master
median:0.657ADU
fixed thresholds:0.2...1.0ADU
N_data plotted:87
[click plot for closeup]
Plot 4
scores: 4|
data source:fors2_bias
(QC1 database)
dataset:noise_master
median:0.725ADU
fixed thresholds:0.2...1.0ADU
N_data plotted:87
[click plot for closeup]
This plot

For the master an expected master bias noise is calculated from the ron of the raw data, divided by the square root of the number of frames that were used to create the master bias. Then the population standard deviation of the pixel values of the master bias is determined, excluding all pixels that differ more than three times the expected ron from the median level of the master bias. Details can be found in the plot tutorial.

DB Name: fors2_bias
TPL.ID: img_cal_bias
Fits header keyword: QC.MBIAS.NOISE
QC1 DB parameter: noise_master

 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up