OMEGACAM trending system: HISTORY report
HELP USERS-GUIDE MORE
Last update: 2017-04-15T03:30:31 (UT)
same group: BIAS_level BIAS_readnoise RON_select BIAS_select BIAS_hotpixels
close window history: 2011 2012 2013 2014 2015 2016   2017   2018   2019 2020 2021 2022 2023 2024   HEALTH
  ... ... ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
    OmegaCAM temperature probe keywords no longer exist. Until fixed, temp_AVG can be ignored.
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 2.06 ADU VAL1.5,2.5 214 qc_ron this | last_yr | all BIAS readnoise level averaged over all 32 detectors (ADU)
2 QC1DB MEDIAN 2.04 ADU none  6847 qc_ron this | last_yr | all BIAS readnoise level for each individual detector (ADU)
3 QC1DB MEDIAN 0.115 ADU VAL0.,0.5 214 qc_ron this | last_yr | all BIAS readnoise level rms averaged over all 32 detectors (ADU)
4 LOCAL none   none none  6781 det_id n/a score per detector:
GREEN = within the thresholds
4 LOCAL none   none none  67 det_id n/a score per detector:
RED = beyond the thresholds
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
data source:omegacam_readnoise
(QC1 database)
dataset:qc_ron
median:2.06ADU
fixed thresholds:1.5...2.5ADU
N_data plotted:214
[click on plot for closeup]
Plot 2
data source:omegacam_readnoise
(QC1 database)
dataset:qc_ron
median:2.04ADU
thresholds:none 
N_data plotted:6847
[click on plot for closeup]
Plot 3
data source:omegacam_readnoise
(QC1 database)
dataset:qc_ron
median:0.115ADU
fixed thresholds:0.0...0.5ADU
N_data plotted:214
[click on plot for closeup]
Plot 4
data source:local data source
dataset:
(numbers below apply to this dataset)
det_id
average:none
thresholds:none
N_data plotted:6781
[click on plot for closeup]
This plot

Trending plots for the bias read-out noise.

QC parameters are derived from each of 32 detectors of OmegaCAM BIAS frames.

DB Name: omegacam_readnoise
Fits header: QC.READNOISE
QC1 DB: qc_ron / qc_temp_mosaic_plate

Description:
The bias read-out noise as measured from the pipeline product frame READNOISE_ADU

Setup:
BIN.X / BIN.Y / READ.MODE / READ.SPEED = 1/1/normal/normal

These Health Check plots describe:
Fig. 1 readnoise_AVG: The bias read-out noise averaged over all 32 detectors (in ADU). This plot is sensitive to any changes common to all detectors.
Fig. 2 readnoise_ALL: The bias read-out noise for each individual detector (in ADU). This plot is sensitive changes in individual detectors.
Fig. 3 readnoise_RMS: The rms of the bias read-out noise averaged over all 32 detectors (in ADU).
Fig. 4 score_outliers: Scores of the bias readnoise as measured for each of the 32 detectors.
Green = within the defined thresholds.
Red = beyond the defined thresholds.


OmegaCAM detector plane layout (detector_ID locations):
OmegaCAM detector layout

 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up