[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SINFONI trending system: FULL report
HELP USERS-GUIDE MORE
Last update: 2019-06-28T05:11:51 (UT)
  | now: 2024-05-05T21:21:29 (UT) muc04 QC pipeline: sinfo-3.2.5 (installed 2019-05-09)
close window history: 2009 2010 2011 2012 2013 2014 2015 2016 2017 2018   2019   HEALTH FULL
  ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 oQC1DB MEDIAN 25.3 ADU VAL2.5,300.0 1044 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture in the J-band at all camera scales (ADU)
This is NOT scored
1 QC1DB none   ADU none  768 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: J-band/250 mas scale
1 QC1DB none   ADU none  179 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: J-band/100 mas scale
1 QC1DB none   ADU none  97 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: J-band/25 mas scale

2 oQC1DB MEDIAN 20.1 ADU VAL2.5,300.0 1252 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture in the H-band at all camera scales (ADU)
This is NOT scored
2 QC1DB none   ADU none  835 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: H-band/250 mas scale
2 QC1DB none   ADU none  369 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: H-band/100 mas scale
2 QC1DB none   ADU none  48 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: H-band/25 mas scale

3 oQC1DB MEDIAN 13.9 ADU VAL2.5,300.0 2965 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture in the K-band at all camera scales (ADU)
This is NOT scored
3 QC1DB none   ADU none  1511 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: K-band/250 mas scale
3 QC1DB none   ADU none  1164 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: K-band/100 mas scale
3 QC1DB none   ADU none  290 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: K-band/25 mas scale

4 oQC1DB MEDIAN 16.9 ADU VAL2.5,300.0 2223 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture in the H+K-band at all camera scales (ADU)
This is NOT scored
4 QC1DB none   ADU none  1065 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: H+K-band/250 mas scale
4 QC1DB none   ADU none  512 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: H+K-band/100 mas scale
4 QC1DB none   ADU none  646 qc_net_flux all (full time range) mean flux per pixel within a 3 sigma aperture: H+K-band/25 mas scale

*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:sinfoni_std
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_net_fluxo
median:25.3ADU
fixed thresholds:2.5...300.0ADU
N_data plotted:1044
[click plot for closeup]
Plot 2
scores: not implemented
data source:sinfoni_std
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_net_fluxo
median:20.1ADU
fixed thresholds:2.5...300.0ADU
N_data plotted:1252
[click plot for closeup]
Plot 3
scores: not implemented
data source:sinfoni_std
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_net_fluxo
median:13.9ADU
fixed thresholds:2.5...300.0ADU
N_data plotted:2965
[click plot for closeup]
Plot 4
scores: not implemented
data source:sinfoni_std
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_net_fluxo
median:16.9ADU
fixed thresholds:2.5...300.0ADU
N_data plotted:2223
[click plot for closeup]
This plot

These are the trending plots for the SINFONI QC parameters derived from the telluric STD stars.

Telluric standard star monitoring has been implemented beginning 2009-05-01.

The intent is to monitor the MINIMUM flux of the STD star to insure that it has an adequate S/N.

Flux levels should be above 5 ADU/pixel within a 3 sigma aperture.

Find more information about SINFONI STD calibration data here.

Other QC parameters are also trended here.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up