[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SINFONI trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2015-08-30T18:01:16 (UT)
  | now: 2015-08-30T21:24:47 (UT) muc04 QC pipeline: sinfo-2.6.5 (installed 2015-08-05)
same group: lamp_eff   lamp_badpix lamp_off lamp_fpn1 lamp_fpn2 vignetting_check_sl1 vignetting_check_sl32
close window history: 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014   2015   HEALTH FULL
  ... ... ... ... ... ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 oQC1DB MEDIAN 0.283 ADU VAL0.23,0.355 25 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): J-band ALL camera scales
(this FULL data set is scored)
1 QC1DB none   ADU none  15 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): J-band/250 mas scale
1 QC1DB none   ADU none  7 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): J-band/100 mas scale
1 QC1DB none   ADU none  3 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): J-band/ 25 mas scale

2 oQC1DB MEDIAN 0.17 ADU VAL0.12,0.255 23 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): H-band ALL camera scales
(this FULL data set is scored)
2 QC1DB none   ADU none  14 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): H-band/250 mas scale
2 QC1DB none   ADU none  7 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): H-band/100 mas scale
2 QC1DB none   ADU none  2 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): H-band/ 25 mas scale

3 oQC1DB MEDIAN 0.157 ADU VAL0.10,0.235 113 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): K-band ALL camera scales
(this FULL data set is scored)
3 QC1DB none   ADU none  59 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): K-band at the 250 mas camera scale
3 QC1DB none   ADU none  38 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): K-band at the 100 mas camera scale
3 QC1DB none   ADU none  16 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): K-band at the 25 mas camera scale

4 oQC1DB MEDIAN 0.181 ADU VAL0.14,0.275 100 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): H+K-band ALL camera scales
(this FULL data set is scored)
4 QC1DB none   ADU none  33 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): H+K-band at the 250 mas camera scale
4 QC1DB none   ADU none  30 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): H+K-band at the 100 mas camera scale
4 QC1DB none   ADU none  37 qc_lflat_fpn1 this | last_yr | all fixed pattern noise (region 1): H+K-band at the 25 mas camera scale
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 2|
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_lflat_fpn1o
median:0.283ADU
fixed thresholds:0.23...0.355ADU
N_data plotted:25
[click on plot for closeup]
Plot 2
scores: 3|
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_lflat_fpn1o
median:0.17ADU
fixed thresholds:0.12...0.255ADU
N_data plotted:23
[click on plot for closeup]
Plot 3
scores: 9|
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_lflat_fpn1o
median:0.157ADU
fixed thresholds:0.1...0.235ADU
N_data plotted:113
[click on plot for closeup]
Plot 4
scores: 3|
data source:sinfoni_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_lflat_fpn1o
median:0.181ADU
fixed thresholds:0.14...0.275ADU
N_data plotted:100
[click on plot for closeup]
This plot

The FPN is calculated over 2 windows:

FPN1: over the central region of the detector(X: 512-1536 Y: 512-1536
FPN2: over the area of a single slitlet (X: 1360-1400 Y: 1000-1200)

The area where FPN2 is computed was shifted by 10 pixels after the intervention which took place second part of 2013.

Remark: For H+K the area for FPN2 contains some lines, which causes a correlation with humidity  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up