[page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
SINFONI trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2014-11-27T05:05:07 (UT)
  | now: 2014-11-27T06:40:30 (UT) muc04 QC pipeline: sinfo-2.5.4 (installed 2014-10-20)
same group: std_flux std_peak_AO std_strehl_AO std_persistence std_peak_all std_strehl_all
close window history: 2009 2010 2011 2012 2013   2014   HEALTH FULL
  ... ... ... ...
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
telluric standard star monitoring has been implemented beginning 2009-05-01. The intent is to monitor the MINIMUM flux of the STD to insure that it has an adequate S/N. Flux on the peak should be around 500 counts which corresponds to ~ 15 ADU/pixel within a 3 sigma aperture for the 25mas scale with AO.
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 oQC1DB no MEDIAN 30.1 ADU VAL2.5,300.0 136 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture in the J-band at all camera scales (ADU)
(this FULL data set is scored)
1 QC1DB no none   ADU none  101 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: J-band/250 mas scale
1 QC1DB no none   ADU none  28 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: J-band/100 mas scale
1 QC1DB no none   ADU none  7 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: J-band/25 mas scale

2 oQC1DB no MEDIAN 22.5 ADU VAL2.5,300.0 174 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture in the H-band at all camera scales (ADU)
(this FULL data set is scored)
2 QC1DB no none   ADU none  128 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: H-band/250 mas scale
2 QC1DB no none   ADU none  35 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: H-band/100 mas scale
2 QC1DB no none   ADU none  11 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: H-band/25 mas scale

3 oQC1DB no MEDIAN 15 ADU VAL2.5,300.0 199 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture in the K-band at all camera scales (ADU)
(this FULL data set is scored)
3 QC1DB no none   ADU none  120 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: K-band/250 mas scale
3 QC1DB no none   ADU none  62 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: K-band/100 mas scale
3 QC1DB no none   ADU none  17 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: K-band/25 mas scale

4 oQC1DB no MEDIAN 21.8 ADU VAL2.5,300.0 271 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture in the H+K-band at all camera scales (ADU)
(this FULL data set is scored)
4 QC1DB no none   ADU none  54 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: H+K-band/250 mas scale
4 QC1DB no none   ADU none  116 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: H+K-band/100 mas scale
4 QC1DB no none   ADU none  101 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: H+K-band/25 mas scale

5 oQC1DB no MEDIAN 27.8 ADU VAL10,300 136 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: any Band, 25mas scale
5 QC1DB no none   ADU none  7 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: J-band/25 mas scale
5 QC1DB no none   ADU none  11 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: H-band/25 mas scale
5 QC1DB no none   ADU none  17 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: K-band/25 mas scale
5 QC1DB no none   ADU none  101 qc_net_flux this | last_yr | all mean flux per pixel within a 3 sigma aperture: H+K-band/25 mas scale
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: 1|
data source:sinfoni_std
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_net_fluxo
median:30.1ADU
fixed thresholds:2.5...300.0ADU
N_data plotted:136
[click on plot for closeup]
Plot 2
scores: 19|
data source:sinfoni_std
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_net_fluxo
median:22.5ADU
fixed thresholds:2.5...300.0ADU
N_data plotted:174
[click on plot for closeup]
Plot 3
scores: 9|
data source:sinfoni_std
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_net_fluxo
median:15ADU
fixed thresholds:2.5...300.0ADU
N_data plotted:199
[click on plot for closeup]
Plot 4
scores: 5|
data source:sinfoni_std
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_net_fluxo
median:21.8ADU
fixed thresholds:2.5...300.0ADU
N_data plotted:271
[click on plot for closeup]
Plot 5
scores: not implemented
data source:sinfoni_std
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_net_fluxo
median:27.8ADU
fixed thresholds:10...300ADU
N_data plotted:136
[click on plot for closeup]
This plot

These are the trending plots for the SINFONI QC parameters derived from the telluric STD stars.

Telluric standard star monitoring has been implemented beginning 2009-05-01.

The intent is to monitor the MINIMUM flux of the STD star to insure that it has an adequate S/N.

Flux levels should be above 5 ADU/pixel within a 3 sigma aperture.

Find more information about SINFONI STD calibration data here.

Other QC parameters are also trended here.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up